11
RF Input VSWR ( at tuned frequency )
< 1.5 : 1
< 1.8 : 1
Attenuator setting 0 dB < 1.8 : 1
Attenuator setting 10 dB < 1.8 : 1
< 1.4 : 1
Attenuator setting 20 dB < 1.6 : 1
10 MHz to 3.0 GHz, Nominal
2.5 GHz to 3.0 GHz, Typical
10 MHz to 2.5 GHz, Typical
10 MHz to 3.0 GHz, Typical
100 kHz to 10 MHz, Nominal
100 kHz to 10 MHz, Nominal
Spurious response
Second harmonic distortion :
Third - order intermodulation :
( third order intercept )
Input related spurious :
Inherent residual response :
Mixer level = – 40 dBm, frequency 50
MHz
Third-order intermodulation products, 2 x – 20 dBm,
reference level –10 dBm, center frequency 300 MHz,
– 40 dBm signal at input mixer,
carry offest > 1 MHz.
Input terminated and 0 dB RF attenuation,
preamplifier off, reference level – 30 dBm,
f > 30 MHz, RBW
<
10 kHz
frequency separation 200 kHz
RF attenuation = 0 dB
RF preamplifier = OFF
Sweep
Sweep time
Range :
Sweep mode :
Trigger source :
Trigger slope :
Trigger delay :
Range
Resolution
10 ms to 1000 s
6
µ
s to 200 s
Span
1 kHz
Span
= 0 Hz ( zero span )
Continuous, single
Free run, video, external
Selectable positive or negative edge
6
µs to 200 s
6
µs
Front panel input / output
RF input
Connector and impedance :
Nominal
10 MHz reference / External trigger input
Reference input frequency :
Reference input amplitude :
Trigger voltage :
Connector and output impedance :
BNC female, 50
Type - N female, 50
5 V TTL level
0 to + 10 dBm
10 MHz
Nominal
Nominal
<
– 70 dBc
<
– 70 dBc
<
– 88 dBm
+
10 dBm, Typical
Connectivity
USB host :
USB device :
USB Type-A female,compatible with
USB 2.0 full speed
USB Type-mini AB female
LAN :
RJ-45, 10 Base-T
Compatible with USB 2.0 full speed