General specifications, Environmental, Remote operation – Atec Agilent-8566B User Manual
Page 13
![background image](/manuals/318714/13/background.png)
General Specifications
.
Environmental
Temperature
Operation 0” to 55” C
Storage
-40” to 75” c
Increased internal temperatures may result if the
rear panel air filters are not cleaned regularly.
Altitude
Operation
4,572 m (15,000 ft)
Storage
15,240 m (50,000 ft)
Power Requirements
50 to 60 Hz; 100,200, 120,220,
or 240 V
-10%); approximately 650 VA (40 VA in
standby). 400 Hz operation with Option 400.
Humidity
Operation
Type tested to 95% relative humidity,
25” to 40” C, except as noted in electrical
specifications.
Storage
5% to 90% relative humidity,
to
C
Conducted and radiated interference is within the
requirements of
Part 7 RE02 and
CEO3 (Air Force), and
Publication 11;
VDE 0871 and FTZ
Warm-Up Time
Operation
Requires 30 minute warm-up from cold
start, 0” to 55” C. Internal temperature
equilibrium is reached after 2-hour
warm-up at stable outside temperature.
Frequency Reference (typical)
Frequency reference aging rate attained
after 24 hour warm-up from cold start at
25” C. Frequency is within 1 x
of final
stabilized frequency within 30 minutes.
Weight
Total, net
50 kg (112 lb)
RF section, net
29 kg (65 lb)
IF display section, net
21 kg (47 lb)
RF section, shipping
35 kg (78 lb)
IF display section, shipping
27 kg (60 lb)
Dimensions
(Allow 100 mm, 4 inch
clearance at rear panel
for
cables.
marked with
denote with handles.)
‘457.2 mm (18 in.)
mm (24.7 in)
598.5 mm (23.56 in)
5mm (16 75
BACK
558.8 mm (22 in)
SIDE
Remote Operation
The standard HP 8566B operates on the
Packard Interface Bus (HP-IB). All analyzer control
settings (with the exception of VIDEO TRIGGER
LEVEL, FOCUS, ALIGN, INTENSITY, FREQ ZERO,
AMPTD CAL, and LINE power) are remotely
programmable. Function values, marker frequency/
A/B traces may be output; CRT
labels and graphics may be input.
LCL Returns analyzer to local control, if not locked out by
controller.
Service Request
SHIFT r calls an HP-IB request for service.
Interface Functions
T6,
SRI,
PPO,
Cl, C2,
C3, C28, E2
Options
All specifications for options are identical to standard
HP 85668 except as noted.
400 Hz Power Line Frequency Operation (Option 400)
Power Line Related Sidebands
(center frequency from 100 Hz to 5.8
Offset from Carrier
Sideband Level
2
-55
2
to 5.5
-65
Power Requirements
Line frequency
400 Hz
line frequency
(50 to 60 Hz operation for
servicing only)
Line voltage
100 to 120 v
-10%)
Operating Temperature Range
400 Hz
0” to 55” c
50 Hz to 60 Hz (service only,
not for extended periods) 0” to
C
Fig.7. Instrument dimensions with and without handles