Frequency specifi cations – Atec Agilent-E4443A-45A-40A-47A-46A-48A User Manual
Page 4
4
Frequency Specifi cations
Frequency range
E4443A
(DC coupled)
3 Hz to 6.7 GHz
(AC coupled)
20 MHz to 6.7 GHz
E4445A
(DC coupled)
3 Hz to 13.2 GHz
(AC coupled)
20 MHz to 13.2 GHz
E4440A
(DC coupled)
3 Hz to 26.5 GHz
1
(AC coupled)
20 MHz to 26.5 GHz
1
E4447A
(DC coupled)
3 Hz to 42.98 GHz
E4446A
(DC coupled)
3 Hz to 44 GHz
1
E4448A
(DC coupled)
3 Hz to 50 GHz
1
1. 325 GHz with external mixers
Band
Harmonic mixing mode (N)
0
1–
3 Hz to 3 GHz
1
1–
2.85 GHz to 6.6 GHz
2
2–
6.2 GHz to 13.2 GHz
3
4–
12.8 GHz to 19.2 GHz
4
4–
18.7 GHz to 26.8 GHz
5
4+
26.4 GHz to 31.15 GHz
6
8–
31.0 GHz to 50.0 GHz
Frequency reference
Accuracy
±[(time since last adjustment x aging rate) + temperature stability + calibration accuracy]
Aging rate
± 1 x 10
–7
/ year
Temperature stability
20 °C to 30 °C ±1 x 10
–8
0 °C to 55 °C ±5 x 10
–8
Achievable initial calibration accuracy
±7 x 10
–8
Example frequency reference accuracy
1 year after last adjustment
= ±(1 x 1 x 10
–7
+ 1 x 10
–8
+ 7 x 10
–8
)
= ±1.8 x 10
–7
Frequency readout accuracy (start, stop, center, marker)
± (marker frequency x frequency reference accuracy + 0.25% x span + 5% x RBW + 2 Hz + 0.5 x horizontal resolution*)
* Horizontal resolution is span/(sweep points – 1)
Marker frequency counter
Accuracy
±(marker frequency x frequency reference accuracy + 0.100 Hz)
Delta counter accuracy
±(delta frequency x frequency reference accuracy + 0.141 Hz)
Counter resolution
0.001 Hz
Frequency span (FFT and swept mode)
Range
0 Hz (zero span), 10 Hz to maximum frequency of model
Resolution 2
Hz
Accuracy
±[0.2% x span + span / (sweep points – 1)]