Power suite measurement specifi cations – Atec Agilent-E4443A-45A-40A-47A-46A-48A User Manual
Page 16
![background image](/manuals/318692/16/background.png)
16
Channel power
Amplitude accuracy, W-CDMA or IS95
(20 to 30 °C, mixer level < –20 dBm)
±0.68 dB ( ±0.18 dB typical)
Occupied bandwidth
Frequency accuracy
±[span/600] nominal
Adjacent channel power
Accuracy, W-CDMA (ACLR)
(at specifi c mixer levels and ACLR ranges)
Adjacent Alternate
MS
±0.12 dB ±0.17 dB
BTS
±0.22 dB ±0.22 dB
Dynamic range (typical)
Without noise correction
–74.5 dB –82 dB
With noise correction
–81 dB –88 dB
Offset channel pairs measured ACP
speed (fast method). Data measurement
and transfer time 30 ms nominal (0.2 dB
standard deviation)
1 to 6
Multi-carrier power and ACP
ACPR dynamic range, W-CDMA
(5 MHz offset, RRC weighted, 3.84 MHz
noise bandwidth)
Two carriers
–70 dB nominal
Four carriers
–66 dB nominal
With noise correction
–76 dB nominal
ACPR accuracy
(two carriers, 5 MHz offset, –48 dBc ACPR)
±0.38 dB nominal
Multiple number of carriers measured
Up to 12
Power statistics CCDF
Histogram resolution
0.1 dB
Harmonic distortion
Maximum harmonic number
10th
Results
Fundamental power (dBm), relative harmonics power (dBc), total harmonic distortion in %
Intermod (TOI)
Measure the third-order products and intercepts from two tones
Burst power
Methods
Power above threshold, power within burst width
Results
Single burst output power, average output power, maximum power, minimum power
within burst, burst width
Power Suite Measurement Specifi cations