beautypg.com

Amplitude specifi cations, Continued) – Atec Agilent-E4443A-45A-40A-47A-46A-48A User Manual

Page 11

background image

11

Amplitude Specifi cations

(continued)

Input voltage standing wave ratio (VSWR) (≥ 8 dB input attenuation)

50 MHz to 3 GHz

< 1.2:1 nominal

3 GHz to 18 GHz

< 1.6:1 nominal

18 GHz to 26.5 GHz

< 1.9:1 nominal

26.5 GHz to 50 GHz

< 1.57:1 nominal

Preamp on (50 MHz to 3 GHz) (≥ 10 dB
attenuation)

< 1.2:1 nominal

Preamp on (Option E444xA-110)
(> 10 dB input attenuation)
E4443A/E4445A/E4440A

200 MHz to 6.6 GHz

< 1.4:1

6.6 to 13.2 GHz

< 1.7:1

13.2 to 19.2 GHz

< 1.5:1

19.2 to 26.5 GHz

< 1.8:1

E4447A/E4446A/E4448A

200 MHz to 6.6 GHz

< 1.2:1

6.6 to 13.2 GHz

< 1.4:1

13.2 to 19.2 GHz

< 1.3:1

19.2 to 31 GHz

< 1.5:1

31 to 50 GHz

< 1.7:1

Resolution bandwidth switching uncertainty (referenced to 30 kHz RBW)

1 Hz to 1 MHz RBW

±0.03 dB

1.1 MHz to 3 MHz RBW

±0.05 dB

4, 5, 6, 8 MHz RBW

±1.0 dB

Reference level

Range

Log scale –170 dBm to +30 dBm in 0.01 dB steps
Linear scale 707 pV to 7.07 V in 0.1% steps

Accuracy 0

dB

Display scale switching uncertainty

Switching between linear and log

0 dB

Log scale/div switching

0 dB

Display scale fi delity

≤ –20 dBm input mixer level

±0.07 dB total

–20 dBm < mixer level ≤ –10 dBm

±0.13 dB total

Spurious response (mixer level = –40 dBm)

General spurious

100 Hz ≤ f < 10 MHz from carrier

(–73 + 20 log N) dBc

f ≥ 10 MHz from carier

(–80 + 20 log N) dBc, (–90 + 20 log N) dBc typical

See frequency range for N