Amplitude specifications
Output level
–10 dBm to +1 dBm
10 dBm to +2.8 dBm (typical)
Resolution 0.1
dB
Accuracy
Vernier
±0.20 dB, ±0.5 dBm max (25 °C ±10 °C)
Absolute
±0.75
dB
Level
flatness
±2.0
dB
Effective source match
1.92:1 (nominal)
Total absolute accuracy
±3.25 dB
Spurious output (at +1 dBm output power)
Harmonic
spurious
–25
dBc
Non-harmonic spurious
300 kHz to 2.0 GHz
–27 dBc
2.0 GHz to 2.9 GHz
–23 dBc
LO feedthrough
–16 dBm (3.9 GHz to 6.8 GHz)
Residuals (RF-power-off)
–78 dBm (300 kHz to 2.9 GHz
Dynamic range
TG feedthrough
1
300 kHz to 1 MHz
–95 dBm
1 MHz to 2.7 GHz
–115 dBm
2.7 GHz to 2.9 GHz
–110 dBm
Dynamic range
2
300 kHz to 1 MHz
96 dB
1 MHz to 2.7 GHz
116 dB
2.7 GHz to 2.9 GHz
111 dB
Power sweep
10 dB range, 0.1 dB resolution
Inputs/outputs
RF output (front panel)
Type-N female, 50 Ω (nominal)
Maximum safe reverse level
+ 30 dBm, ±30 Vdc
External ALC input (rear panel)
BNC female use with negative detector
10
1. Leakage measured with maximum power into 50 Ω, with 50 Ω loads on the
TG output and RF input.