Fig. 12
Placing an All-weather
Case on location

Logging in the Field

Unattended measurements require protection against weather
and unauthorised access, and the All-weather Case Type 3535-A
fills the bill. Lightweight and robust, it will house a Type 2250-H
Hand-held Analyzer and supply it with battery power for up to
90 hours of operation. A wide range of optional accessories are
available to tailor the system capability to the logging task.
Outdoor Microphone Type 4952 is shown in Fig. 12. When the
system is left unattended, measurements and status can be
monitored remotely with a wide range of options (see the
Connectivity section).

For details, please refer to

Product Data BP 2251

.

Fig. 13
Automatic Charge
Injection Calibration at
beginning and end of a
measurement

The Exclude marker
shows that calibration
is not part of the
measurement

Charge Injection Calibration (CIC)

Measurement integrity is vital in any measurement application,
particularly if the system is operating unattended. CIC provides a means
of regularly auto-checking the measurement system for 10 s to ensure
the continued and accurate operation of the system.

With the Logging software, CIC can be set to run at the beginning and
end of measurement.

With the Enhanced Logging Software BZ-7225, CIC can be programmed
to run at up to four user-specified times in each 24 hour period. It may
also be activated manually.

Fig. 14
Possible metadata
entry types

Shown are six user-
defined metadata
items and a
selection menu for the
Location item – this
takes just four
taps on the screen

Metadata the Easy Way

Metadata are supplementary information about your measurement to
help in archiving, retrieval and post-processing. Examples of metadata
are the file name, date and time, setup and annotations made by
the operator.

The names and types of up to 10 text strings can be further defined. The
type may be text you enter, text from a pick list you define, a number you
enter or an index number that is automatically incremented or
decremented when the measurement is saved.

Metadata are available in all applications and may be used for sorting
measurements in Measurement Partner Suite BZ-5503.

Enhanced Logging Software – BZ-7225

Fig. 15
Display showing part of
a report profile with
10 min resolution

With the optional Enhanced Logging Software enabled, your analyzer is
optimised for extended logging periods. It has the functionality of both
Logging and Frequency Analysis software, but in addition it will:

Measure continuously, saving data to SD memory cards
or USB devices

Reboot automatically and resume operation in case of
power failure

Save data in manageable portions (every 24 hours),
selectable for download

Make periodic reports, i.e., log all measurement data at a
preset report period

Measure L

dn

, L

den

, L

day

, L

evening

and L

night

120005