beautypg.com

Compliance with standards – Atec Bruel-Kjaer-2250-2270 User Manual

Page 12

background image

12

Compliance with Standards

Hand-held Analyzer Type 2250/2270 Platform

Specifications apply to Type 2250/2270 fitted with Microphone Type 4189
and Microphone Preamplifier ZC-0032

Microphone

SUPPLIED MICROPHONE
Type 4189:
Prepolarized Free-field ½

Microphone

or
Type 4190: Free-field ½

Microphone

Nominal Open-circuit Sensitivity: 50 mV/Pa (corresponding to
26 dB re 1 V/Pa) ± 1.5 dB
Capacitance: 14 pF (at 250 Hz)

MICROPHONE PREAMPLIFIER ZC-0032
Nominal Preamplifier Attenuation:
0.25 dB
Connector: 10-pin LEMO
Extension Cables: Up to 100 m in length between the microphone
preamplifier and Type 2250/2270, without degradation of the
specifications
Accessory Detection: Windscreen UA-1650 can be automatically
detected when fitted over ZC-0032

MICROPHONE POLARIZATION VOLTAGE
Selectable between 0 V and 200 V

SELF-GENERATED NOISE LEVEL
Typical values at 23°C for nominal microphone open-circuit sensitivity:

Interface

KEYBOARD
Pushbuttons:
11 keys with backlight, optimised for measurement control
and screen navigation

ON-OFF BUTTON
Function:
Press 1 s to turn on; press 1 s to enter standby; press for more
than 5 s to switch off

STATUS INDICATORS
LEDs:
Red, amber and green

DISPLAY
Type:
Transflective back-lit colour touch screen 240

 320 dot matrix

Colour Schemes: Five different – optimised for different usage
scenarios (day, night, etc.)
Backlight: Adjustable level and on-time

USER INTERFACE
Measurement Control:
Using pushbuttons on keyboard
Setup and Display of Results: Using stylus on touch screen or
pushbuttons on keyboard
Lock: Keyboard and touch screen can be locked and unlocked

USB INTERFACE
USB 2.0 OTG Micro AB and USB 2.0 Standard A sockets

MODEM INTERFACE
Connection to Internet through GPRS/EDGE/HSPA modem connected
through:
• the USB Standard A Socket
Supports DynDNS for automatic update of IP address of host name

PRINTER INTERFACE
PCL printers, Mobile Pro Spectrum thermal printer or Seiko DPU
S245/S445 thermal printers can be connected to USB socket

MICROPHONE FOR COMMENTARY
Microphone, which utilises Automatic Gain Control (AGC), is
incorporated in underside of analyzer. Used to create voice annotations
for attaching to measurements

CAMERA (TYPE 2270 ONLY)
Camera with fixed focus and automatic exposure is incorporated in
underside of analyzer.
Used to create image annotations for attaching to measurements
Image Size:
• 2048

 1536 pixels

Viewfinder Size: 212

 160 pixels

Format: jpg with exif information

Inputs/Outputs

SECURE DIGITAL SOCKET
• 2

 SD sockets

Connect SD and SDHC memory cards

LAN INTERFACE SOCKET
• Connector: RJ45 Auto-MDIX
• Speed: 100 Mbps
• Protocol: TCP/IP

CE-mark indicates compliance with the EMC Directive and Low Voltage Directive.
C-Tick mark indicates compliance with the EMC requirements of Australia and New Zealand.

Safety

EN/IEC 61010–1, ANSI/UL 61010–1 and CSA C22.2 No.1010.1: Safety requirements for electrical equipment for
measurement, control and laboratory use.

EMC Emission

EN/IEC 61000–6–3: Generic emission standard for residential, commercial and light industrial environments.
CISPR 22: Radio disturbance characteristics of information technology equipment. Class B Limits.
FCC Rules, Part 15: Complies with the limits for a Class B digital device.
IEC 61672–1, IEC 61260, IEC 60651 and IEC 60804: Instrumentation standards.
Complies with Canadian standard ICES–001

EMC Immunity

EN/IEC 61000–6–2: Generic standard – Immunity for industrial environments.
EN/IEC 61326: Electrical equipment for measurement, control and laboratory use – EMC requirements.
IEC 61672–1, IEC 61260, IEC 60651 and IEC 60804: Instrumentation standards

Weighting

Microphone

Electrical

Total

“A”

14.6 dB

12.4 dB

16.6 dB

“B”

13.4 dB

11.5 dB

15.6 dB

“C”

13.5 dB

12.9 dB

16.2 dB

“Z” 5 Hz–20 kHz

15.3 dB

18.3 dB

20.1 dB

“Z” 3 Hz–20 kHz

15.3 dB

25.5 dB

25.9 dB