beautypg.com

Atec Agilent-35670A User Manual

Page 8

background image

8

Computed order tracking – Option 1D0

Maximum order x Maximum RPM

(

——————————————

)

60

Online (real time)

1 channel mode 25,600 Hz
2 channel mode 12,800 Hz
4 channel mode 6,400 Hz

Capture playback

1 channel mode 102,400 Hz
2 channel mode 51,200 Hz
4 channel mode 25,600 Hz

Number of orders ≤ 200

5 ≤ RPM ≤ 491,519

(Maximum useable RPM is limited by resolution, tach pulse
rate,pulses/revolution and average mode settings.)
Delta order

1/128 to 1/1

Resolution

≤ 400

(Maximum order)/(Delta order)
Maximum RPM ramp rate 1000 RPM/second real-time

(typical)

1000 - 10,000 RPM run up
Maximum order

10

Delta order

0.1

RPM step

30 (1 channel)
60 (2 channel)
120 (4 channel)

Order track amplitude
accuracy

±1 dB (typical)

Real time octave analysis – Option 1D1

Standards
Conforms to ANSI Standard S1.11 - 1986, Order 3, Type 1-D,
extended and optional frequency ranges
Conforms to IEC 651-1979 Type 0 Impulse, and ANSI S1.4

1 second stable average
Single tone at band center:

≤ ± 0.20 dB

Readings are taken from the linear total power spectrum
bin. It is derived from sum of each filter.
1/3-octave dynamic range

> 80 dB (typical) per
ANSI S1.11-1986

2 second stable average
Total power limited by input noise level

Frequency ranges (at centers)
Online (real time):

Single channel

2 channel

4 channel

1/1 octave

0.063 - 16 kHz

0.063 - 8 kHz

0.063 - 4 kHz

1/3 octave

0.08 - 40 kHz

0.08 - 20 kHz

0.08 - 10 kHz

1/12 octave

0.0997 -

12.338 kHz

0.0997 -

6.169 kHz

0.0997 -

3.084 kHz

Capture playback
1/1 octave

0.063 - 16 kHz

0.063 - 16 kHz

0.063 - 16 kHz

1/3 octave

0.08 - 31.5 kHz

0.08 - 31.5 kHz

0.08 - 31.5 kHz

1/12 octave

0.0997 -

49.35 kHz

0.0997 -

49.35 kHz

0.0997 -

49.35 kHz

One to 12 octaves can be measured and displayed.
1/1-, 1/3-, and 1/12-octave true center fre quen cies related by the
formula: f(i+1)/f(i) = 2^(1/n); n=1, 3, or 12; where 1000 Hz is
the reference for 1/1, 1/3 octave, and 1000*2^(1/24) Hz is the
reference for 1/12 octave. The marker returns the ANSI standard
preferred frequencies.

Swept sine measurements – Option 1D2

Dynamic range

130 dB

Tested with 11 dBVrms source level at: 100 mSec integration

Curve fi t/synthesis – Option 1D3

20 Poles/20 zeroes curve filter frequency response
synthesis pole/zero, pole residue & polynomical format

Arbitrary waveform source – Option 1D4

Amplitude range

AC: ±5 V peak*
DC: ±10 V*

* Vac

pk

+ |Vdc| ≤ 10 V

Record length

# of points = 2.56 x lines of
resolution, or # of complex
points = 1.28 x lines of resolution

DAC resolution
0.2828 Vpk to 5 Vpk

2.5 mV

0 Vpk to 0.2828 Vpk

0.25 mV