Agilent 35670A Specifi cations

Instrument specifications apply after 15 minutes warm-up
and within 2 hours of the last self-calibration. When the
internal cooling fan has been turned OFF, specifications
apply within 5 minutes of the last self-cal i bra tion. All speci-
fications are with 400 line frequency resolution and with
anti-alias filters enabled unless stated otherwise.

Frequency

Maximum range**
1 channel mode

102.4 kHz,
51.2 kHz (opt AY6*)

2 channel mode

51.2 kHz

4 channel mode (Option AY6 only) 25.6 kHz
Spans
1 channel mode

195.3 mHz to 102.4 kHz

2 channel mode

97.7 mHz to 51.2 kHz

4 channel mode (Option AY6 only) 97.7 mHz to 25.6 kHz
Minimimum resolution
1 channel mode

122 µHz (1600 line
display)

2 channel mode

61 µHz (1600 line
display)

4 channel mode (Option AY6 only) 122 µHz (800 line

display)

Maximum real-time bandwidth

FFT span for continuous data acquistion)
(Preset, fast averaging)

1 channel mode

25.6 kHz

2 channel mode

12.8 kHz

4 channel mode (Option AY6 only) 6.4 kHz
Measurement rate

(Typical) (Preset, fast averaging)

1 channel mode

≥ 70 averages/sec

2 channel mode

≥ 33 averages/sec

4 channel mode (Option AY6 only) ≥ 15 averages/sec
Display update rate
Typical (Preset, fast average off)

≥ 5 updates/Sec

Maximum

≥ 9 updates/Sec

(Preset, fast average off, single channel, single display,
undisplayed trace displays set to data registers)

Accuracy

±30 ppm (.003%)
Single channel ampltude
Absolute amplitude accuracy (FFT)
(A combination of full scale accuracy, full scale flatness,
and amplitude linearity.)
±2.92% (0.25 dB) of reading
±0.025% of full scale
FFT full scale accuracy at 1 kHz (0 dBfs)
±0.15 dB (1.74%)
FFT full scale flatness (0 dBfs) relative to 1 kHz
±0.2 dB (2.33%)
FFT amplitude linearity at 1 kHz measured on +27 dBVrms
range with time avg, 0 to -80 dBfs
±0.58% (0.05 dB) of reading
±0.025% of full scale
Amplitude resolution
(16 bits less 2 dB over-range) with averaging 0.0019% of
full scale (typical)
Residual DC response (FFT mode)
Frequency display (excludes A-weight filter)
<-30 dBfs or 0.5 mVdc
FFT dynamic range
Spurious free dynamic range
(Includes spurs, harmonic distortion, intermodulation
distortion, alias products). Excludes alias responses at
extremes of span.
Source impedence = 50 Ω.
800 line display.
90 dB typical (<-80 dBfs)

* Option AY6 single channel maximum range extends to 102.4 kHz

without anti-alias fi lter protection.

** Show all lines mode allows display of up to 131.1, 65.5 and 32.7

kHz respectively. Amplitudes accuracy is unspecifi ed and not alias