beautypg.com

Internal modulation source, Lf out (internal modulation source), External modulation inputs – Atec Agilent-E4437B User Manual

Page 9: Simultaneous modulation

background image

9

Internal modulation source

(Provides FM,

ΦM, and AM modulation signals and LF out)

Waveforms

sine, square, ramp, triangle,
pulse, noise

Rate range
Sine

0.1 Hz to 50 kHz

Square, ramp, triangle

0.1 Hz to 10 kHz

Resolution

0.1 Hz

Pulse only

4 µs

Frequency accuracy

0.005%, typical

Swept sine mode (frequency, phase continuous)
Operating modes

Triggered or continuous sweeps

Frequency range

0.1 Hz to 50 kHz

Sweep time

1 ms to 65 sec

Resolution

1 ms

Dual sinewave mode
Frequency range

0.1 Hz to 50 kHz

Amplitude ratio

0 to 100%

Amplitude ratio resolution

0.1%

LF out (internal modulation source)

Amplitude

0 to 3 V

peak

into 50

Output impedance

<1

External modulation inputs

Modulation types
Ext 1

FM,

ΦM, AM, and burst envelope

Ext 2

FM,

ΦM, AM, and pulse

High/Low Indicator (100 Hz to 10 MHz BW, AC coupled inputs
only) Activated when input level error exceeds 3% (nominal)

Simultaneous modulation

All modulation types may be simultaneously enabled, except: FM
with FM; AM with burst envelope; Wideband AM with I/Q. AM,
FM, and FM can sum simultaneous inputs from any two sources
(INT, EXT 1, and EXT 2.) Any given source (INT, EXT 1, or EXT 2)
may only be routed to one activated modulation type.