Spectral purity – Atec Agilent-E4437B User Manual
Page 5
5
Resolution
0.02 dB
Attenuator hold level range
Standard
Option UNB
250 kHz to 1 GHz
23 dB
27 dB
>1 to 3 GHz
20 dB
26 dB
>3 to 4 GHz
17 dB
23 dB
Level accuracy (dB)
1
Output power
+7 to –120 dBm
(+10 to –120 dBm, –120 to
Freq range
Option UNB)
–127 dBm
<–127 dBm
250 kHz to 2 GHz ±0.5
±0.5
±1.5
2 to 3 GHz
±0.9
±0.9
±2.5
3 to 4 GHz
±0.9
±0.9 (±1.5,
±2.5
Option UNB)
Typical level accuracy
Amplitude switching speed
Without power search
<30 ms, typical
When using power search
<300 ms, typical
Reverse power protection
2
250 kHz to 2 GHz
50 watts
>2000 to 4 GHz
25 watts
Max DC voltage
50 V
SWR (typical)
Standard
Option UNB
250 kHz to 1 GHz
<1.5:1
<1.3:1
1 to 2 GHz
<1.4:1
<1.3:1
2 to 3 GHz
<1.3:1
<1.4:1
3 to 4 GHz
<1.5:1
<1.5:1
Output impedance
50
Ω
Spectral purity
SSB phase noise
3
(at 20 kHz offset)
ESG-A and
ESG-AP and
ESG-D Series
ESG-DP Series
at 500 MHz (<–120 dBc/Hz)
<–134 dBc/Hz, (<–138 dBc/Hz)
at 1 GHz
(<–116 dBc/Hz)
<–130 dBc/Hz, (<–134 dBc/Hz)
at 2 GHz
(<–110 dBc/Hz)
<–123 dBc/Hz, (<–127 dBc/Hz)
at 3 GHz
(<–104 dBc/Hz)
<–120 dBc/Hz, (<–124 dBc/Hz)
at 4 GHz
(<–104 dBc/Hz)
<–118 dBc/Hz, (<–122 dBc/Hz)
Residual FM
4
(CW mode, 0.3 to 3 kHz BW, CCITT, rms)
ESG-AP and ESG-DP series
ESG-A and ESG-D series Phase noise mode 2 Harmonics ( ≤+4 dBm (≤+7.5 dBm, Option UNB) output level) <–30 dBc (typical below 1 GHz) Nonharmonics (<+7 dBm (<+10 dBm, Option UNB) output level) 5 ESG-A andESG-AP and 6 ESG-DP series 7 >3 kHz >10 kHz >3 kHz >10 kHz offset offset 3 offset offset 3 250 kHz to 250 MHz <–65 dBc (<–75 dBc) <–65 dBc (<–75 dBc) 250 MHz to 500 MHz <–65 dBc (<–75 dBc) <–80 dBc <–80 dBc 500 MHz to 1 GHz (<–65 dBc) (<–75 dBc) <–80 dBc <–80 dBc 1 to 2 GHz (<–59 dBc) (<–69 dBc) <–74 dBc <–74 dBc >2 GHz (<–53 dBc) (<–63 dBc) <–68 dBc <–68 dBc Subharmonics ESG-A and ESG-AP and ESG-D series ESG-DP series ≤1 GHz None None >1 GHz (<–40 dBc) None Characteristic ESG-A and ESG-D series SSB phase Frequency (MHz) Level error (dBm) Option IE5 Standard PN2 PN2 PN1 PN1 1. For 23 °C ±5 °C. Accuracy degrades by 0.02 dB/°C over the full temperature range and by 0.3 dB above +7 dBm (degraded by 0.5 dB above +10 dBm with Option UNB). Level accuracy specification maintained only with return to calibration. 2. The reverse power protection circuitry triggers at nominally 1 watt. For non-constant amplitude digital formats, unspecified spur levels occur up to the second harmonic of the baseband rates. 7. Specifications apply for CW mode only.
Phase noise mode 1
ESG-D series
noise at 1 GHz (phase noise modes 1 and 2)
3. Parentheses denote typical performance.
4. Refer to frequency bands on page 4 to compute specifications.
5. Performance is typical for spurs at frequencies above the maximum operating frequency of the instrument. Performance typically is –60 dBc between 225 and 249.999 MHz.
6. Specifications apply for FM deviations <100 kHz and are not valid for FM.