Ordering information, Supplemental characteristics for all model numbers – Atec Agilent-N3300A User Manual
Page 8
![background image](/manuals/318509/8/background.png)
8
Ordering Information
N3300A & N3301A mainframes ship
with product reference CD-ROM
including: VXIplug&play drivers, user’s
guide, programming guide, and
application notes
Module options:
Opt. A6J: Certificate of calibration
Opt. UK6: Commercial calibration
Opt. UJ1: 8 mm screw terminal
connector
Chassis options:
Opt. 0L1: Printed user’s guide and
programming guide
Note: no service manual available for
this product
Accessories
1CM020A* (N3300A)
Rackmount flange kit 88.1 mm H (2U)
– four brackets (4U total); 1.75 inch
hole spacing
1CP012A* (N3300A)
Rackmount flange and handle kit
88.1 mm H (2U) – four brackets (4U
total); front handles
1CM001A* (N3301A)
Rackmount flange kit 177.0 mm H (4U)
– one bracket, one half-module
bracket
1CM034A* (N3301A)
Rackmount flange kit 177.0 mm H
(4U) – two flange brackets and lock
link kit for mounting two units side by
side. Equivalent to 1CM023A plus
p/n 5061-9694 lock link kit.
E3663AC
Support rails for Agilent rack cabinets
* Support rails or slides are required.
Supplemental
Characteristics for All
Model Numbers
Command processing time:
Average time for the output voltage to
change after getting a GPIB command
is 3 ms for discrete commands, 1 ms
for list commands
List dwell characteristics:
Range: 0 - 10 s
Resolution: 1 ms
Accuracy: 5 ms
Transient generator:
Frequency range: 0.25 Hz - 10 kHz
Frequency accuracy: 0.5%
Duty cycle range:
3 to 97% (0.25 Hz - 1 kHz);
6 to 94% (1 kHz - 10 kHz)
Duty cycle accuracy: 1%
Pulse width: 50 μs ± 1% to
4 seconds ± 1%
Analog programming bandwidth:
10 kHz rms (-3 db frequency)
Analog programming voltage:
Voltage: 0 - 10 V
Current: 0 - 10 V
Analog monitor ports:
Voltage: 0 - 10 V
Current: 0 - 10 V
DC floating voltage:
Output terminals
can be floated up to ± 300 VDC from
chassis ground
Remote sensing:
5 V DC between sense and load input
Digital/Trigger inputs:
Vil = 0.9 V max at Iil = -1 mA
Vih - 3.15 V min (pull-up resistor on
input)
Digital/Trigger outputs:
Vol = 0.72 V max at Iol = 1 mA
Voh = 4.4 V min at Ioh = -20 μA
GPIB interface capabilities:
SH1, AH1, T6, L4, SR1, RL1 DT1, CD1
Software driver:
VXIplug&play
Regulatory compliance:
UL 61010B-1, IEC 61010-1/EN 61010-1,
CSA C22.2 No. 1010.1
Net weight:
N3300A: 13.2 kg (29 lb);
N3301A: 7.3 kg (16 lb);
N3302A, N3303A, N3304A,
N3304A-J08 or N3307A: 2.7 kg (6 lb);
N3305A or N3306A: 4.6 kg (10 lb)
Shipping weight:
N3300A: 17 kg (38 lb);
N3301A: 9.1 kg (20 lb)
N3302A, N3303A, N3304A,
N3304A-J08 or N3307A: 4.1 kg (9 lb)
N3305A or N3306A: 6.8 kg (15 lb)
Calibration interval:
One year for
modules, mainframes do not require
calibration
Warranty:
One year