Specifications – Atec Agilent-N3300A User Manual
Page 4
4
Specifications
Table A-1 lists the specifications for the different load models. Specifications
indicate warranted performance in the 25 °C ± 5 °C region of the operating
temperature range. Specifications apply to normal and transient modes unless
otherwise noted.
Table A-1
N3302A
N3303A
N3304A
N3304-J01
Special
order option
N3305A
N3306A
N3307A
Input ratings
Current
0–30 A
0–10 A
0–60 A
0–60 A
0–60 A
0–120 A
0–30 A
Voltage
0–60 V
0–240 V
0–60 V
0–80 V
0–150 V
0–60 V
0–150 V
Maximum power @ 40 °C
1
150 W
250 W
300 W
300 W
500 W
600 W
250 W
Input characteristic
Operating contour
Derated current detail
Voltage
Full
scale
Full
scale
0
Max. power contour
Current
Voltage
Full
scale
Full
scale
3
2
1
0
All specifications apply
Slew rate limitations apply
(see Table A-2)
Current
N3302A
N3303A
N3304A
N3304-J01
Special
order option
N3305A
N3306A
N3307A
Specified current @ low voltage operation
2.0 V
30 A
10 A
60 A
60 A
60 A
120 A
30 A
1.5 V
22.5 A
7.5 A
45 A
45 A
45 A
90 A
22.5 A
1.0 V
15 A
5 A
30 A
30 A
30 A
60 A
15 A
0.5 V
7.5 A
2.5 A
15 A
15 A
15 A
30 A
7.5 A
0 V
0 A
0 A
0 A
0 A
0 A
0 A
0 A
1. Maximum continuous power available is derated linearly from 100% of maximum at 40 °C, to 75% of maximum at 55 °C.
Table A-1 states that maximum current is available down to 2 volts. Typically,
however under normal operating conditions, the load can sink the maximum
current down to the following voltages:
N3302A
N3303A
N3304A
N3304-J01
Special
order option
N3305A
N3306A
N3307A
Typical minimum operating voltage @ full scale current
1.2 V
1.2 V
1.2 V
1.2 V
1.4 V
1.4 V
1.4 V