Table 2. Measurement Throughput Summary

Typical Cycle Time

1, 2

(ms) for Measurement Completion

Description

Typical Performance
(time/point in millisecond)

Number of Points

CW 10 GHz (no band crossings), 801 points

Trigger Mode

Hardware

IF Bandwidth

600
kHz

100
kHz

10
kHz

1
kHz

RF = MXG, N5183A opt. UNZ, Fast switching
LO = MXG, N5183A opt. UNZ, Fast switching

0.070

0.075

0.185

1.00

RF = MXG, N5183A opt. UNZ, Fast switching

LO = N5264A opt. 108

3

0.070

0.075

0.185

1.00

RF = MXG, N5183A opt. UNZ, Fast switching
LO = PSG

0.350

0.350

0.450

0.250

RF = MXG, N5183A opt. UNZ, Fast switching
LO = 83623B

0.900

0.900

1.00

1.800

Typical Cycle Time

1, 2

(ms) for Measurement Completion (Cont.)

Description

Typical Performance
(time/point in millisecond)

Number of Points

801 1601

Trigger Mode

Hardware

Sensitivity(dBm)

4

Start 2 GHz, Stop 18 GHz, 1 MHz IF bandwidth (with band crossings)

RF = MXG, N5183A opt. UNZ, Fast switching
LO = MXG, N5183A opt. UNZ, Fast switching

0.580

0.580

-90.5 dBm, 2 – 3 GHz
- 94.5 dBm, 3 – 12.5 GHz
- 83 dBm, 12.5 – 18 GHz

RF = MXG, N5183A opt. UNZ, Fast switching
LO = N5264A opt. 108

3

0.580

0.580

-85.5 dBm, 2 – 3 GHz
- 90.5 dBm, 3 – 12.5 GHz
- 81 dBm, 12.5 – 18 GHz

Start 2 GHz, Stop 18 GHz, 600 kHz IF bandwidth (with band crossings)

RF = MXG, N5183A opt. UNZ, Fast switching
LO = MXG, N5183A opt. UNZ, Fast switching

0.580

0.580

-92.5 dBm, 2 – 3 GHz
- 96.5 dBm, 3 – 12.5 GHz
- 85 dBm, 12.5 – 18 GHz

RF = MXG, N5183A opt. UNZ, Fast switching
LO = N5264A opt. 108

3

0.580

0.580

-85.5 dBm, 2 – 3 GHz
- 92.5 dBm, 3 – 12.5 GHz
- 83 dBm, 12.5 – 18 GHz