beautypg.com

Vna master functional specifications, Definitions, Frequency – Atec Anritsu-VNA Series User Manual

Page 3: Transmission dynamic range, Test port power, Sweep speed, High-level noise (s, Or s, Temperature stability (s, Short, 23 ºc ± 5 ºc, typical)

background image

3

VNA Master Functional Specifications

Definitions

All specifications and characteristics apply under the following conditions, unless otherwise stated:

• After 30 minutes of warm-up time, when the instrument is in VNA Mode and left in the ON state.

• Temperature range is 23 ºC ± 5 ºC.

• All specifications apply when using internal reference.

• All specifications subject to change without notice. Please visit www.anritsu.com for most current data sheet.

• Typical performance is the measured performance of an average unit.

• Recommended calibration cycle is 12 months.

Frequency

VNA Frequency Range:

MS2026/36C: 5 kHz to 6 GHz

MS2027/37C: 5 kHz to 15 GHz

MS2028/38C: 5 kHz to 20 GHz

Frequency Accuracy:

±1.5 ppm

Frequency Resolution:

1 Hz to 375 MHz, 10 Hz to 6 GHz, and 100 Hz to 20 GHz

Frequency Range

High Port Power

dBm, typical)

Low Port Power

(dBm, typical)

5 kHz to ≤ 3 GHz

+3

–25

3 GHz to ≤ 6 GHz

–3

–25

6 GHz to ≤ 20 GHz

–3

–15

Frequency Range

Dynamic Range

(dB)

5 kHz to ≤ 2 MHz

85

2 MHz to ≤ 3 GHz

100

3 GHz to ≤ 6 GHz

90

6 GHz to ≤ 20 GHz

85

Transmission Dynamic Range

The transmission dynamic range (the difference between

test port power and noise floor) using 10 Hz IF Bandwidth

and High Port Power is shown in the following table.

Test Port Power

VNA Master supports selection of either High (default) or

Low test port power. Changing power after calibration can

degrade the calibrated performance. Typical power by bands

is shown in the following table.

Sweep Speed

The typical sweep speed for IF Bandwidth of 100 kHz, 1001

data points, and single display is shown in the following

table. The three receiver architecture will simultaneously

collect S

21

and S

11

(or S

12

and S

22

) in a single sweep.

Frequency Range

Sweep Speed

(µsec/point, typical)

5 kHz to 6 GHz

350

6 GHz to 20 GHz

650

Magnitude

Phase

0.004 dB(rms) (5 kHz to 6 GHz)

0.040 deg (5 kHz to 6 GHz)

0.010 dB(rms) (6 GHz to 20 GHz)

0.050 deg (6 GHz to 20 GHz)

High-Level Noise (S

11

or S

22

, Short, Power = High,

IFBW = 200 Hz typical)

Noise Floor (Port Power - Dynamic Range)

Frequency

Noise Floor (dB, typical)

5 kHz to = 2 MHz

Port Power –85

2 MHz to = 3 GHz

Port Power –100

3 GHz to = 6 GHz

Port Power –90

6 GHz to = 20 GHz

Port Power –85

Temperature Stability (S

11

or S

22

, Short, 23 ºC ± 5 ºC, typical)

Magnitude

Phase

0.018 dB/ºC (5 kHz to 10 GHz)

0.160 deg/ºC (5 kHz to 10 GHz)

0.070 dB/ºC (10 GHz to 20 GHz)

0.800 deg/ºC (10 GHz to 20 GHz)

Source Match* (Anritsu 3652A Cal Kit)

Frequency

Match (dB, typical)

5 kHz to 1 GHz

41

1 GHz to 5 GHz

39

5 GHz to 20 GHz

31

Reflection Tracking (S

11

or S

22

)

Frequency

Tracking (dB, typical)

< 3 GHz

0.05

3 to 6 GHz

0.10

6 to 20 GHz

0.20

Transmission Tracking (S

21

or S

12

)

Frequency

Tracking (dB, typical)

< 3 GHz

0.02

3 to 6 GHz

0.05

6 to 20 GHz

0.40

Load Match* (Anritsu 3652A Cal Kit)

Frequency

Match (dB, typical)

5 kHz to 1 GHz

37

1 GHz to 15 GHz

34

15 GHz to 20 GHz

30