Spectrum Analyzer Functional Specifications (MS203xC models only) (continued)

Third-Order Intercept (TOI) (–20 dBm tones 100 kHz apart, –20 dBm Ref level, 0 dB input attenuation, preamp Off)
2.4 GHz

+15 dBm

50 MHz to 20 GHz

+20 dBm typical

P1dB
< 4 GHz

+5 dBm typical

4 GHz to 20 GHz

+12 dBm typical

Second Harmonic Distortion
50 MHz

–54 dBc

< 4 GHz

–60 dBc typical

> 4 GHz

–75 dBc typical

VSWR
> 10 dB input attenuation

< 20 GHz

1:5:1 typical

Measurements

Smart Measurements

Field Strength (uses antenna calibration tables to measure dBm/m2 or dBmV/m)

Occupied Bandwidth (measures 99% to 1% power channel of a signal)

Channel Power (measures the total power in a specified bandwidth)

ACPR (adjacent channel power ratio)

C/I (carrier-to-interference ratio)

Emission Mask (recall limit lines as emission mask)

Setup Parameters
Frequency

Center/Start/Stop, Span, Frequency Step, Signal Standard, Channel #

Amplitude

Reference Level (RL), Scale, Attenuation Auto/Level, RL Offset, Pre-Amp On/Off, Detection

Span

Span, Span Up/Down (1-2-5), Full Span, Zero Span, Last Span

Bandwidth

RBW, Auto RBW, VBW, Auto VBW, RBW/VBW, Span/RBW

File

Save, Recall, Copy, Delete, Directory Management

Save/Recall

Setups, Measurements, Limit Lines, Screen Shots Jpeg (save only), Save-on-Event

Save-on-Event

Crossing Limit Line, Sweep Complete, Save-then-Stop, Clear All

Delete

Selected File, All Measurements, All Mode Files, All Content

Directory Management

Sort Method (Name/Type/Date), Ascend/Descend, Internal/USB, Copy

Application Options

Impedance (50 Ω, 75 Ω, Other)

Sweep Functions
Sweep

Single/Continuous, Manual Trigger, Reset, Detection, Minimum Sweep Time, Trigger Type

Sweep Mode

Fast, Performance, No FFT, Burst Detect

Detection

Peak, RMS/Avg, Negative, Sample, Quasi-peak

Triggers

Free Run, External, Video, Delay, Level, Slope, Hysteresis, Holdoff, Force Trigger Once

Trace Functions
Traces

Up to three Traces (A, B, C), View/Blank, Write/Hold, Trace A/B/C Operations

Trace A Operations

Normal, Max Hold, Min Hold, Average, # of Averages, (always the live trace)

Trace B Operations

A→B, B↔C, Max Hold, Min Hold

Trace C Operations

A→C, B↔C, Max Hold, Min Hold, A – B→C, B – A→C, Relative Reference (dB), Scale

Marker Functions

Markers

Markers 1-6 each with a Delta Marker, or Marker 1 Reference with Six Delta Markers,

Marker Table (On/Off/Large), All Markers Off

Marker Types

Style (Fixed/Tracking), Noise Marker, Frequency Counter Marker

Marker Auto-Position

Peak Search, Next Peak (Right/Left), Peak Threshold %, Set Marker to Channel,

Marker Frequency to Center, Delta Marker to Span, Marker to Reference Level

Marker Table

1-6 markers frequency and amplitude plus delta markers frequency offset and amplitude

Limit Line Functions
Limit Lines

Upper/Lower, On/Off, Edit, Move, Envelope, Advanced, Limit Alarm, Default Limit

Limit Line Edit

Frequency, Amplitude, Add Point, Add Vertical, Delete Point, Next Point Left/Right

Limit Line Move

To Current Center Frequency, By dB or Hz, To Marker 1, Offset from Marker 1

Limit Line Envelope

Create Envelope, Update Amplitude, Number of Points (41), Offset, Shape Square/Slope

Limit Line Advanced

Type (Absolute/Relative), Mirror, Save/Recall