ZERO SET (digital settability of zero):
±0.07% of full scale on lowest range.
Decrease by a factor of 10 for each higher range.
Supplemental Characteristics:
ZERO DRIFT OF METER:
±0.03% of full scale/°C on lowest range. Decrease by a
factor of 10 for each higher range.
NOISE (at constant temperature, peak change over any one-
minute interval for the 11722A or 11792A sensor modules):
0.4% of full scale on range 1 (lowest range).
0.13% of full scale on range 2.
0.013% of full scale on range 3.
0.0013% of full scale on range 4.
0.00013% of full scale on range 5.
ZERO DRIFT OF SENSORS (1 hour, at constant temperature after
24-hour warm-up):
±0.1% of full scale on lowest range for 11722A
and 11792A sensor modules.
RF POWER RANGES OF AGILENT 8902A MEASURING RECEIVER
WITH 11722A AND 11792A SENSOR MODULES:
–20 dBm to –10 dBm (10 µW to 100 µW), range 1.
–10 dBm to 0 dBm (100 µW to 1 mW), range 2.
0 dBm to +10 dBm (1 mW to 10 mW), range 3.
+10 dBm to +20 dBm (10 mW to 100 mW), range 4.
+20 dBm to +30 dBm (100 mW to 1 W), range 5.
RESPONSE TIME (0 to 99% of reading):
<10 seconds, range 1.
<1 second, range 2.
<100 milliseconds, ranges 3 through 5.
DISPLAYED UNITS:
Watts, dBm, dB
relative
, %
relative
, volts, mV, µV, dB V, dB mV, dB µV.
INTERNAL NON-VOLATILE CAL-FACTOR TABLES
(user-modifiable using special functions):
Maximum number of cal factor/frequency entries:
Table #1 (primary): 16 pairs plus Reference Cal Factor.
Table #2 (frequency offset): 22 pairs plus Reference Cal Factor.
Maximum Allowed Frequency Entry: 42 GHz.
Frequency Entry Resolution: 50 kHz.
Cal Factor Range: 40 to 120%.
Cal Factor Resolution: 0.1%.
Power Reference
POWER OUTPUT:
1.00 mW. Factory set to ±0.7%, traceable to the U.S. National
Bureau of Standards.
ACCURACY: ±1.2% worst case (±0.9% rss) for one year (0 °C to
55 °C).
Supplemental Characteristics:
FREQUENCY: 50 MHz nominal.
SWR: 1.05 nominal.
FRONT PANEL CONNECTOR: N-type female.
Tuned RF Level
POWER RANGE: –127 dBm to 0 dBm, using IF synchronous detector
(200 Hz BW).
–100 dBm to 0 dBm, using IF average detector (30 kHz BW).
POWER RANGE (Using 11792A Sensor Module):
For IF Synchronous Detector:
+10 dBm to –117 dBm, 2.5 MHz
≤
f
c
≤
1300 MHz.
+5 dBm to –105 dBm, 1300 MHz
≤
f
c
≤
12.4 GHz.
+5 dBm to –100 dBm, 12.4 GHz
≤
f
c
≤
18.0 GHz.
+5 dBm to –95 dBm, 18.0 GHz
≤
f
c
≤
26.5 GHz.
For IF Average Detector:
+10 dBm to –90 dBm, 2.5 MHz
≤
f
c
≤
1300 MHz.
+5 dBm to –80 dBm, 1300 MHz
≤
f
c
≤
12.4 GHz.
+5 dBm to –75 dBm, 12.4 GHz
≤
f
c
≤
18.0 GHz.
+5 dBm to –70 dBm, 18.0 GHz
≤
f
c
≤
26.5 GHz.
1.9 Special Function degrades Tuned RF Level minimum
sensitivity by 10 dB.
FREQUENCY RANGE:
2.5 MHz to 1300 MHz.
2.5 MHz to 26.5 GHz.
DISPLAYED RESOLUTION
11
:
4 digits in watts or volts mode.
0.01 dB or 0.001 dB in dBm or dB
relative
mode.
4 digits in watts or volts mode.
0.01 dB in dBm or dB
relative
mode.
RELATIVE MEASUREMENT ACCURACY (at constant temperature
and after RF range calibration is completed)
12
:
Detector linearity + IF range-to-range error + RF range-to-range
error + frequency drift error + noise error ± 1 digit.
Detector linearity + mixer linearity + IF range-to-range error + RF
range-to-range error + frequency drift error + noise error ± 1 digit.
11. The 8902A fundamental Tuned RF Level measurement units are volts. Further
internal processing is done on this number to display all other units.
12. Tuned RF Level accuracy will be affected by residual FM of the source-under-test.
If the residual FM
peak
is >50 Hz measured over a 30 second period in a 3 kHz BW.
Tuned RF Level measurements should be made using the IF average detector (30
kHz BW) by using Special Function 4.4. The Tuned RF Level measurement sensi-
tivity when using the IF average detector is –100 dBm.