AGILENT 11812A VERIFICATION KIT
SPECIFICATIONS
FREQUENCY: 30 MHz.
11812A ACCURACY: ±(0.003 dB + 0.003 dB/10 dB step).
OPTION 050 WORST CASE CUMULATIVE TUNED RF LEVEL
ACCURACY VERIFIED WITH 11812A:
±0.010 dB/10 dB step (0 to –100 dBm).
±0.050 dB/10 dB step (–100 to –120 dBm).
±0.015 dB ± 1 digit.
TEMPERATURE:
Operation: 15 °C to 30 °C.
Storage: –55 °C to 74 °C.
AGILENT 8902A REAR PANEL
INPUTS/OUTPUTS
Supplemental Characteristics:
FM OUTPUT: 10 k
Ω
impedance, –9 V to 6 V into an open circuit,
~6 V/MHz, dc coupled, 16 kHz bandwidth (one pole).
AM OUTPUT: 10 k
Ω
impedance, –4 V to 0 V into an open circuit,
~8 mV/%, dc coupled, 16 kHz bandwidth (one pole).
RECORDER OUTPUT: DC voltage proportional to the measured
results, 1 k
Ω
impedance, 0 V to 4 V for each resolution range into
an open circuit.
IF OUTPUT: 50
Ω
impedance, 150 kHz to 2.5 MHz, –27 dBm to –3
dBm.
10 MHz REFERENCE OUTPUT: 50
Ω
impedance, TTL levels (0 V to
>2.2 V into an open circuit). Available only with Option 002
1x10
–9
/day internal reference.
10 MHz REFERENCE INPUT
17:
>500
Ω
impedance, 0.5 V
peak-to-peak
minimum input level.
LO INPUT (Option 003): 50
Ω
impedance, ~1.27 MHz to 1301.5
MHz, 0 dBm nominal.
RF SWITCH REMOTE CONTROL OUTPUT: Provides output signals
necessary to remotely control either an Agilent 33311B,C Option
011 or an 8761A RF switch.
FREQUENCY OFFSET MODE REMOTE CONTROL OUTPUT: TTL
high output if in frequency offset mode (Special Function 27.1 or
27.3) with an external LO frequency >0, TTL low output for all
other cases.
17. External reference accuracy affects accuracy of all measurements.