Model

SU−221

SU−241

SU−261

SU−641

SU−661

System

Balanced Temperature Control system (BTC system)

T

e

m

p

. p

e

rf

o

rm

a

n

c

e

*1

Temp. range

−20 to +150℃

(−4 to +302℉)

−40 to +150℃

(−40 to +302℉)

−60 to +150℃

(−76 to +302℉)

−40 to +150℃

(−40 to +302℉)

−60 to +150℃

(−76 to +302℉)

Temp. fl uctuation

±0.3℃ (−20 to +100℃)
±0.5℃ (+100.1 to +150℃)

±0.3℃ (−40 to +100℃)
±0.5℃ (+100.1 to +150℃)

±0.3℃ (−60 to +100℃)
±0.5℃ (+100.1 to +150℃)

±0.3℃ (−40 to +100℃)
±0.5℃ (+100.1 to +150℃)

±0.3℃ (−60 to +100℃)
±0.5℃ (+100.1 to +150℃)

Temp. gradient /
Temp. variation in space

2.5℃ (−20 to +100℃)
4.0℃ (+100.1 to +150℃)

2.5℃ (−40 to +100℃)
4.0℃ (+100.1 to +150℃)

2.5℃ (−60 to +100℃)
4.0℃ (+100.1 to +150℃)

2.5℃ (−40 to +100℃)
4.0℃ (+100.1 to +150℃)

2.5℃ (−60 to +100℃)
4.0℃ (+100.1 to +150℃)

Temp.
rate of
change

Heat up rate

3.2℃ /min.

3.2℃ /min.

3.2℃ /min.

2.9℃ /min.

2.9℃ /min.

Pull down rate

2.1℃ /min.

2.1℃ /min.

2.1℃ /min.

1.7℃ /min.

1.7℃ /min.

Heat up time

From −20 to +150℃

within 55 min.

From −40 to +150℃

within 60 min.

From −60 to +150℃

within 70 min.

From −40 to +150℃

within 70 min.

From −60 to +150℃

within 80 min.

Temp. extreme
achievement time
Pull down time

From +20 to −20℃

within 20 min.

From +20 to −40℃

within 50 min.

From +20 to −60℃

within 70 min.

From +20 to −40℃

within 60 min.

From +20 to−60℃

within 90 min.

Lowest attainable temp.

−20℃

−40℃

−60℃

−40℃

−60℃

Construc

tion

Heater

Nichrome strip wire heater

Ref

ri

g

erat

io

n

un

it System

Mechanical single-stage refrigeration system

Mechanical cascade refrigeration system

Cooler

Plate fi n cooler

Refrigerator

Hermetically sealed compressor, Air-cooled condenser, Expansion mechanism: Capillary tube system

Refrigerator capacity

400W

[Unit 1: 400W ×1, unit 2: 400W ×1]

Refrigerant

R404A

R23, R404A

Capacity

22.5 L

64 L

Chamber total load
resistance

20 kg

Inside dimensions
mm (inch)

*2

W300×H300×D250

(W11.81×H11.81×D9.84)

W400×H400×D400

(W15.75×H15.75×D15.75)

Outside dimensions
mm (inch)

*2

W440×H560×D695

(W17.32×H24.8×D27.36)

W440×H620×D785

(W17.32×H24.41×D30.91)

W540×H660×D920

(W21.26×H25.98×D36.22)

Weight

78 kg (73 for 100V type)

100 kg

123 kg

Utilit

y

requirements

Allowable ambient conditions

+5 to +35℃ (+41 to +95℉)

P

o

we

r s

u

p

p

ly

*3

100V AC 1φ50/60Hz

11.0 A

15.0 A

18.0 A

115V AC 1φ60Hz

10.5 A

14.5 A

200V AC 1φ50/60Hz

10.5 A

220V AC 1φ50/60Hz

7.0 A

9.5 A

10.0 A

220V AC 1φ60Hz

*4

7.0 A

9.5 A

10.0 A

230V AC 1φ50Hz

*4

6.5 A

9.5 A

10.0 A

Noise level

*5

55 dB

59 dB

61 dB

Exhaust heat quantity

3500 kJ/h (836 kcal/h)

4000 kJ/h (955 kcal/h)

5040 kJ/h (1204 kcal/h)

8

SU

*1 The performance values are based on IEC 60068-3-5:2001, JTM K07:2007 for the temperature chamber. Performance fi gures are given for a +23℃ ambient temperature,

65%rh, rated power supply and no specimens inside the test area. However, the lowest attainable temperature is given for a max. ambient temperature of +30 ℃.
Heat-up time is the achieved time from lowest temperature to highest temperature within temperature range.

*2 Excluding protrusions.
*3 At ambient temperature +23℃. For SU-641/661, make sure to check the power capability of your installation beforehand.
*4 Compliance with CE Marking.
*5 Measurements are to be taken in an anechoic room at a height of 1.2m from the fl oor and a distance of 1m from the chamber front panel

(ISO 1996-1:2003 _ A-weighted sound pressure level)

小型環境_e_.indd 8