Model
SU−221
SU−241
SU−261
SU−641
SU−661
System
Balanced Temperature Control system (BTC system)
T
e
m
p
. p
e
rf
o
rm
a
n
c
e
*1
Temp. range
−20 to +150℃
(−4 to +302℉)
−40 to +150℃
(−40 to +302℉)
−60 to +150℃
(−76 to +302℉)
−40 to +150℃
(−40 to +302℉)
−60 to +150℃
(−76 to +302℉)
Temp. fl uctuation
±0.3℃ (−20 to +100℃)
±0.5℃ (+100.1 to +150℃)
±0.3℃ (−40 to +100℃)
±0.5℃ (+100.1 to +150℃)
±0.3℃ (−60 to +100℃)
±0.5℃ (+100.1 to +150℃)
±0.3℃ (−40 to +100℃)
±0.5℃ (+100.1 to +150℃)
±0.3℃ (−60 to +100℃)
±0.5℃ (+100.1 to +150℃)
Temp. gradient /
Temp. variation in space
2.5℃ (−20 to +100℃)
4.0℃ (+100.1 to +150℃)
2.5℃ (−40 to +100℃)
4.0℃ (+100.1 to +150℃)
2.5℃ (−60 to +100℃)
4.0℃ (+100.1 to +150℃)
2.5℃ (−40 to +100℃)
4.0℃ (+100.1 to +150℃)
2.5℃ (−60 to +100℃)
4.0℃ (+100.1 to +150℃)
Temp.
rate of
change
Heat up rate
3.2℃ /min.
3.2℃ /min.
3.2℃ /min.
2.9℃ /min.
2.9℃ /min.
Pull down rate
2.1℃ /min.
2.1℃ /min.
2.1℃ /min.
1.7℃ /min.
1.7℃ /min.
Heat up time
From −20 to +150℃
within 55 min.
From −40 to +150℃
within 60 min.
From −60 to +150℃
within 70 min.
From −40 to +150℃
within 70 min.
From −60 to +150℃
within 80 min.
Temp. extreme
achievement time
Pull down time
From +20 to −20℃
within 20 min.
From +20 to −40℃
within 50 min.
From +20 to −60℃
within 70 min.
From +20 to −40℃
within 60 min.
From +20 to−60℃
within 90 min.
Lowest attainable temp.
−20℃
−40℃
−60℃
−40℃
−60℃
Construc
tion
Heater
Nichrome strip wire heater
Ref
ri
g
erat
io
n
un
it System
Mechanical single-stage refrigeration system
Mechanical cascade refrigeration system
Cooler
Plate fi n cooler
Refrigerator
Hermetically sealed compressor, Air-cooled condenser, Expansion mechanism: Capillary tube system
Refrigerator capacity
400W
[Unit 1: 400W ×1, unit 2: 400W ×1]
Refrigerant
R404A
R23, R404A
Capacity
22.5 L
64 L
Chamber total load
resistance
20 kg
Inside dimensions
mm (inch)
*2
W300×H300×D250
(W11.81×H11.81×D9.84)
W400×H400×D400
(W15.75×H15.75×D15.75)
Outside dimensions
mm (inch)
*2
W440×H560×D695
(W17.32×H24.8×D27.36)
W440×H620×D785
(W17.32×H24.41×D30.91)
W540×H660×D920
(W21.26×H25.98×D36.22)
Weight
78 kg (73 for 100V type)
100 kg
123 kg
Utilit
y
requirements
Allowable ambient conditions
+5 to +35℃ (+41 to +95℉)
P
o
we
r s
u
p
p
ly
*3
100V AC 1φ50/60Hz
11.0 A
15.0 A
18.0 A
115V AC 1φ60Hz
10.5 A
14.5 A
—
200V AC 1φ50/60Hz
—
10.5 A
220V AC 1φ50/60Hz
7.0 A
9.5 A
10.0 A
220V AC 1φ60Hz
*4
7.0 A
9.5 A
10.0 A
230V AC 1φ50Hz
*4
6.5 A
9.5 A
10.0 A
Noise level
*5
55 dB
59 dB
61 dB
Exhaust heat quantity
3500 kJ/h (836 kcal/h)
4000 kJ/h (955 kcal/h)
5040 kJ/h (1204 kcal/h)
8
SU
*1 The performance values are based on IEC 60068-3-5:2001, JTM K07:2007 for the temperature chamber. Performance fi gures are given for a +23℃ ambient temperature,
65%rh, rated power supply and no specimens inside the test area. However, the lowest attainable temperature is given for a max. ambient temperature of +30 ℃.
Heat-up time is the achieved time from lowest temperature to highest temperature within temperature range.
*2 Excluding protrusions.
*3 At ambient temperature +23℃. For SU-641/661, make sure to check the power capability of your installation beforehand.
*4 Compliance with CE Marking.
*5 Measurements are to be taken in an anechoic room at a height of 1.2m from the fl oor and a distance of 1m from the chamber front panel
(ISO 1996-1:2003 _ A-weighted sound pressure level)