Options – Atec Espec-SH-641 User Manual
Page 13
12
Effective recording chart width:
100mm
Number of inputs: Temperature 5
Humidity
1
[SRJ14]
Temperature range: −100 to +150℃
Humidity range: 0 to 100%rh (for SH)
[SRJ12]
Temperature range: −50 to +150℃/
Humidity range: 0 to 100%rh (for SH)
Inner door
Glass door provided inside the chamber,
to obser ve t he cond it ions of t he
specimens.
* A wiper is included for SH models.
* This option is not compatible with a viewing
window or right-opening door.
* Used for observation of the specimens inside
the chamber.
SU model
OPTIONS
Wet-bulb temperature detecting terminal
For the test area wet-bulb temperature
detection. A thermocouple type T
(C o p p e r / C o p p e r- N i c k e l ) o r o f
equivalent thermoelectric capacity is
used. Connector included.
* Not available for SU.
[SRJ25]
Temperature range: −100 to +200℃
Effective recording chart width:
100mm
Number of inputs: Temperature 6
Temperature recorder (Portable type)
Temperature & humidity recorder (Portable type)
Thermocouple
Attached to specimens to measure
specimen temperature.
・2m
・4m
・6m
* Thermocouple type T without ball
(Copper / Copper-Nickel)
* Same as accessory items
Paperless recorder (Portable type)
Records temperature of each section
such as the temperature inside the
chamber.
Data saving cycle: 5 sec
Portable type
External recording media:
CF memory card (256MB)
USB port
Language support: ENG, JPN
[Temperature type]
Temperature range: −100 to +200℃
Number of inputs (Initial setting):
Temperature 1
(5 more channels can be turned ON)
[Temperature and humidity type]
Temperature range: − 50 to +150℃
−100 to +150℃
Humidity range: 0 to 100%rh
Number of inputs (Initial setting):
Temperature 1 / Humidity 1
(4 more channels can be turned ON)
Paperless recorder Portable type
W215×H215mm
Viewing window
SH/SU-221, 241, 261 W215×H215mm
SH/SU-641, 661
W215×H315mm
When equipped with a viewing
window, chamber performance
are as follow:
Temperature uniformity :
±1.0℃ (−20 to +100℃)
±1.5℃ (−100.1 to +150℃)
Humidity uniformity : ±5.0%rh
Temp. heat up rate:
Std
rate
+10 min. (or less)
Temp. pull down rate:
Std
rate
+10 min. (or less)
(+20 to −35℃)
Std rate +30 min. (or less)
(+20 to −60℃)
小型環境_e_.indd 12
12.9.10 10:09:20