Specifications – Atec Espec-TSE-11-A User Manual
Page 7
![background image](/manuals/318306/7/background.png)
6
●
Do not use specimens which are explosive or inflammable, or which contain such
substances. To do so could be hazardous, as this may lead to fire or explosion.
●
Do not place corrosive materials in the chamber. If corrosive substances or
humidifying water is used, the life of the unit may be significantly shortened.
●
Do not place life forms or substances that exceed allowable heat generation.
Be sure to read the user’s manual before operation.
CAUTION
DANGER
SPECIFICATIONS
*1 The performance values are under the conditions of a +23℃ ambient temperature, relative humidity of 65%rh, rated voltage, and no specimen. Heat up time
and pull down time are those of single-unit operation of each chamber.
*2 The performance values are based on IEC60068-3-5:2001, JTM K07:2007.
*3 Temperature heat-up/pull-down time account for performance of each temperature chamber.
*4 Excluding protrusions.
*5 Compliance with CE Marking.
*6 At ambient temperature +23℃.
*7 At 1m from front of chamber, 1.2m from floor. (ISO 1996-1:2003 A-weighted sound pressure level) depending on environment
Model
TSE-11-A
System
2-zone transition by vertical transfer of specimen
Performance
*1
Te
s
t a
re
a
High temp. exposure range
+60 to +200℃ (+140 to +392℉)
Low temp. exposure range
−65 to 0℃ (−85 to +32℉)
Temperature fl uctuation
*2
±0.5℃ (±0.9℉)
Hot
chamber
Pre-heat upper limit
+200℃ (+392℉)
Temp. heat-up time
*3
Ambient temp. to +200℃ (+392℉) within 30 min.
Cl
o
d
chamber
Pre-cool lower limit
−80℃ (−112℉)
Temp. pull-down time
*3
Ambient temp. to −80℃ (−112℉) within 90 min.
T
e
m
p
. r
e
c
o
ve
ry
Recovery conditions
・ 2 zones
High temperature exposure: +150℃ (+302℉), 30 min.
Low temperature exposure: −65℃ (−85℉), 30 min.
・ Sensor position: Upstream
・ Specimen: Plastic molded ICs 2 kg
Temp. recovery time
within 5 min.
Construction
Test area
Shelf brackets on 2 levels of fi xed location
Heater
Stripped wire heater
Refrigeration
unit
System
Mechanical cascade refrigeration system
Compressor
Rotary 1.5 kW ×2
Refrigerant
R508A R404A
Condenser
Air-cooled condenser
Cooler
Plate fi n cooler, cold accumulator
Air circulator
Sirocco fan
Fittings
Specimen power supply control terminal, integrating hour meter without reset,
time signal (2), cable port 50 mm, (right side), casters with leveling feer (4), power cable
Test area load resistance
8 kg
Specimen basket load capacity
2kg per basket (equally distributed load)
Inside dimensions (W×H×D)
320Ч148Ч230mm (12.6Ч5.8Ч9 inch)
Test area capacity
10.9 L
Outside dimensions (W×H×D)
*4
680Ч1625Ч1050mm (26.8Ч64Ч41.3 inch)
Weight
approximately 390kg
Allowable ambient conditions
0 to +40℃ (+32 to +104℉)
Power supply
(Voltage fl uctuation: rating ±10%)
200V AC
3φ 3W 50/60Hz
220V AC
3φ 3W 60Hz
380V AC
3φ 4W 50Hz
400/415V AC
3φ 4W 50Hz
400V AC
*5
3φ 4W 50Hz
Maximum load current
26A
25A
17A
17A
17A
Exhaust heat quantity
*6
17,585kJ/h
Noise level
*7
60dB or less