Test standard (tse-11-a compliant) – Atec Espec-TSE-11-A User Manual

Page 6

background image

5

TEST STANDARD (TSE-11-A compliant)

■ The above specification tests include only those tests applicable to TSE-11-A. For further information, please contact us.
* Ambient temperature at exposure temperature and exposure time represents the temperature and time when moving from hot chamber to cold chamber.

Test standard

Exposure temperature

Exposure time

Temp.

recovery time

Number of

cycles

Test starting

point

High temp. Ambient temp.* Low temp.

High/ low temp.

Ambient temp.*

MIL-STD-883H
(Method No.
1010.8)

A

+10

+ 85℃

0

̶

0

− 55℃

−10

more than

10 min.

̶

Specimen

temp within

15 min.

at worst

condition

Minimum

10 cycles

Low or high

temp.

B

+15

+ 125℃

0

C

+15

+ 150℃

0

0

− 65℃

−10

D

+15

+ 200℃

0

F

+15

+ 175℃

0

IEC 60068-2-14
(JIS C 60068-2-14)

+ 70℃ ±2
+ 85℃ ±2
+ 100℃ ±2
+ 125℃ ±2
+ 155℃ ±2
+ 175℃ ±2
+ 200℃ ±2

Ambient

temp.

− 5℃ ±3
− 10℃ ±3
− 25℃ ±3
− 40℃ ±3
− 55℃ ±3
− 65℃ ±3

3

hours

2

hours

1

hour

0.5

hour

3

hours

if not specifi ed

less than 10 sec.

less than 10%

of exposure

time

5 cycles if not

specifi ed

Low temp.

JASO D 001

1

+ 85℃

Ambient

temp.

−40℃

Less than 0.2 kg
1

hour +15 min.

0

0.2∼0.8 kg
2 hours +15 min.

0

0.8∼1.5 kg
3 hours +15 min.

0

More than 1.5 kg
4 hours +15 min.

0

Short

exposure is

recommendable

Upstream of

specimen

within 5 min.

6 cycles

High temp.

2

+ 75℃

3

+120℃

EIAJ ED-2531A

+ 60℃ ±2
+ 65℃ ±2
+ 70℃ ±2
+ 75℃ ±2
+ 80℃ ±2
+ 85℃ ±2
+ 90℃ ±2
+ 95℃ ±2
+ 100℃ ±2

Ambient

temp.

0℃ ±3

− 5℃ ±3
− 10℃ ±3
− 15℃ ±3
− 20℃ ±3
− 25℃ ±3
− 30℃ ±3
− 35℃ ±3
− 40℃ ±3
− 45℃ ±3
− 50℃ ±3

3

hours

2

hours

1

hour

0.5

hour

3

hours

if not specifi ed

less than 10 sec.

less than 10%

of exposure

time

5 or 10 cycles

Low temp.