Test standard (tse-11-a compliant) – Atec Espec-TSE-11-A User Manual
Page 6
5
TEST STANDARD (TSE-11-A compliant)
■ The above specification tests include only those tests applicable to TSE-11-A. For further information, please contact us.
* Ambient temperature at exposure temperature and exposure time represents the temperature and time when moving from hot chamber to cold chamber.
Test standard
Exposure temperature
Exposure time
Temp.
recovery time
Number of
cycles
Test starting
point
High temp. Ambient temp.* Low temp.
High/ low temp.
Ambient temp.*
MIL-STD-883H
(Method No.
1010.8)
A
+10
+ 85℃
0
̶
0
− 55℃
−10
more than
10 min.
̶
Specimen
temp within
15 min.
at worst
condition
Minimum
10 cycles
Low or high
temp.
B
+15
+ 125℃
0
C
+15
+ 150℃
0
0
− 65℃
−10
D
+15
+ 200℃
0
F
+15
+ 175℃
0
IEC 60068-2-14
(JIS C 60068-2-14)
+ 70℃ ±2
+ 85℃ ±2
+ 100℃ ±2
+ 125℃ ±2
+ 155℃ ±2
+ 175℃ ±2
+ 200℃ ±2
Ambient
temp.
− 5℃ ±3
− 10℃ ±3
− 25℃ ±3
− 40℃ ±3
− 55℃ ±3
− 65℃ ±3
3
hours
2
hours
1
hour
0.5
hour
3
hours
if not specifi ed
less than 10 sec.
less than 10%
of exposure
time
5 cycles if not
specifi ed
Low temp.
JASO D 001
1
+ 85℃
Ambient
temp.
−40℃
Less than 0.2 kg
1
hour +15 min.
0
0.2∼0.8 kg
2 hours +15 min.
0
0.8∼1.5 kg
3 hours +15 min.
0
More than 1.5 kg
4 hours +15 min.
0
Short
exposure is
recommendable
Upstream of
specimen
within 5 min.
6 cycles
High temp.
2
+ 75℃
3
+120℃
EIAJ ED-2531A
+ 60℃ ±2
+ 65℃ ±2
+ 70℃ ±2
+ 75℃ ±2
+ 80℃ ±2
+ 85℃ ±2
+ 90℃ ±2
+ 95℃ ±2
+ 100℃ ±2
Ambient
temp.
0℃ ±3
− 5℃ ±3
− 10℃ ±3
− 15℃ ±3
− 20℃ ±3
− 25℃ ±3
− 30℃ ±3
− 35℃ ±3
− 40℃ ±3
− 45℃ ±3
− 50℃ ±3
3
hours
2
hours
1
hour
0.5
hour
3
hours
if not specifi ed
less than 10 sec.
less than 10%
of exposure
time
5 or 10 cycles
Low temp.