Atec EM-Test-UCS500N Series User Manual
Page 6
DATA SHEET > UCS 500N5 > 20131118
TECHNICAL DETAILS
POWER FAIL, DIPS & INTER-
RUPTIONS, VOLTAGE VARIATIONS
POWER FAIL MODULE, PFS/N5
As per EN/IEC 61000-4-11,
IEC/EN 61000-4-29 and
EN 61000-6-1, -6-2
Channel PF1/PF2
AC voltage: max. 300V
AC current: max. 16A
DC voltage: max. 300V
DC current: max. 16A
Frequency
16Hz - 500Hz and DC
Switching time
< 5us into a 100ohm resistive load
Inrush current
> 500A
Protection
Both channels are protected against
short-circuit conditions.
TRIGGER CIRCUIT
Trigger of events
Automatic, manual, external
Synchronization
0° - 360°, resolution 1° (16 - 500Hz)
Repetition rate
10ms - 9,999s
Event duration
20us - 9,999s
OUTPUTS
DUT terminals
L, N and PE
CRO trigger
5V trigger signal for oscilloscope
MEASUREMENTS
DUT voltage
In the LCD display
DUT current
In the LCD display
MON V
Measurement of the DUT voltage;
built-in 100:1 divider
MON I
Measurement of the DUT current;
10mV/A; max. 1,000A
TEST ROUTINES
Quick Start
On-line adjustable parameters,
easy-to-use
Standard Test
routines
As per EN/IEC 61000-4-11
for AC supplies
As per EN/IEC 61000-4-29 for DC
supplies
As per EN 61000-6-1, -6-2
Manual Standard Test routine
User Test routines
Voltage variation, control of an
external variac
Change phase angle after n events
Change event duration after n
events
Inverse mode
50/60Hz magnetic
field
As per EN/IEC 61000-4-8
Test levels 1, 3, 10 and 30A/m with
external current transformer
MC2630
Test levels 100, 300 and 1,000A/m
with external current transformer
MC26100
OPTIONS
V4780
Tapped autotransformer as per IEC
61000-4-11 Ed.2
V4780 S2
Tapped autotransformer as per IEC
61000-4-11 Ed.2 with automatic
change of tap
MV2616
Motorised variac (0 - 250V, 16A)
MS100N
Magnetic field coil, 1m x 1m
MC2630
Current transformer for magnetic
fields up to 30A/m
MC26100
Current transformer for magnetic
fields up to 1,000A/m
CA PFS
Calibration box for inrush current
verification as per IEC 61000-4-11
www.emtest.com
© EM TEST > PAGE 6/8