Ringwave, Telecom surge – Atec EM-Test-UCS500N7 User Manual
Page 7

DATA SHEET > UCS 500N7 > 20120518
TECHNICAL DETAILS
RINGWAVE
RINGWAVE MODULE, RWG/N7 (OPTIONAL)
As per ANSI/IEEE C62.41 and
EN/IEC 61000-4-12
Test voltage
250V - 6,000V ± 10%
Voltage
Wave shape (open circuit)
Rise time (first
peak)
0.5us ± 30%
Oscillation
frequency
100kHz ± 10%
Decaying
Peak 2 to peak 1 = 40 - 110%
Peak 3 to peak 2 = 40 - 80%
Peak 4 to peak 3 = 40 - 80%
Current
Wave shape (short circuit)
Rise time
<= 1.0us
Oscillation
frequency
100kHz ± 10%
Source impedance
12ohm and 30ohm
Peak current
As per selected source impedance
Polarity
Positive, negative
TRIGGER CIRCUIT
Release of pulses
Automatic, manual, external
Synchronization
0° - 360°, resolution 1°
Repetition rate
max. 1Hz (1s - 999s)
OUTPUTS
Direct
Via HV-safety lab connectors
Coupling mode
L, N, PE; line to line and line(s) to
ground
DUT supply
AC: 300V/16A; 50/60Hz
DC: 300V/10A
CRO trigger
5V trigger signal for oscilloscope
TEST ROUTINES
Quick Start
On-line adjustable parameters,
easy-to-use
TELECOM SURGE
TSURGE MODULE, TSURGE7 (OPTIONAL)
Test voltage (o.c.)
250V - 7,000V ± 10%
Energy storage
capacitor
20uF
Polarity
Positive, negative, alternating
Counter
1 - 30,000 or endless, selectable
As per ITU and ETSI
recommendations
Front time
10us ± 30%
Pulse duration
700us ± 20%
As per FCC part 68, Pulse B
Front time
9us ± 30%
Pulse duration
720us ± 20%
Output current
6A - 175A (short circuit)
Rise time
5us ± 30%
Pulse duration
320us ± 20%
As per IEC 61000-4-5
Rise time
6.5us ± 30%
Pulse duration
700us ± 20%
Output current
6A - 175A (short circuit)
Rise time
4us ± 20%
Pulse duration
300us ± 20%
TRIGGER CIRCUIT
Trigger of events
Automatic, manual, external
Repetition rate
max. 0.33Hz (3s - 999s)
OUTPUTS
As per ITU
For 2-wire T1/T2 with 25ohm each
As per FCC part 68
For 2-wire T1/T2 with 25ohm each
As per IEC
61000-4-5
For 4-wire T1/T2/T3/T4 with 100ohm
each
For other requirements special
output configurations are available
OPTIONS
CNV 504S1
Coupling network for 4 telecom lines
as per IEC 61000-4-5
CNV 508S1
Coupling network for 8 telecom lines
as per IEC 61000-4-5
www.emtest.com
© EM TEST > PAGE 7/9