beautypg.com

Telecom surge, General data – Atec EM-Test-UCS500N5 User Manual

Page 7

background image

DATA SHEET > UCS 500N5 > 20120518

TECHNICAL DETAILS

TELECOM SURGE

TSURGE MODULE, TSURGE5 (OPTIONAL)

Test voltage

160V - 5,000V ± 10%

Energy storage
capacitor

20uF

Polarity

Positive, negative, alternating

Counter

1 - 30,000 or endless, selctable

As per ITU and ETSI
recommendations

Front time

10us ± 30%

Pulse duration

700us ± 20%

As per FCC part 68, Pulse B

Front time

9us ± 30%

Pulse duration

720us ± 20%

Output current

4 - 125A (short-circuit)

Rise time

5us ± 30%

Pulse duration

320us ± 20%

As per IEC 61000-4-5

Rise time

6.5us ± 30%

Pulse duration

700us ± 20%

Output current

4A - 125A (short-circuit)

Rise time

4us ± 20%

Pulse duration

300us ± 20%

TRIGGER CIRCUIT

Trigger of events

Automatic, manual, external

Repetition rate

max. 0.5Hz (2s - 999s)

OUTPUTS

As per ITU

For 2-wire T1/T2 with 25ohm each

As per FCC part 68

For 2-wire T1/T2 with 25ohm each

As per IEC
61000-4-5

For 4-wire T1,T2,T3,T4 with 100ohm
each

For other requirements special
output configurations are available

OPTIONS

CNV 504S1

Coupling network for 4 telecom lines
as per IEC 61000-4-5

CNV 508S1

Coupling network for 8 telecom lines
as per IEC 61000-4-5

GENERAL DATA

INTERFACES

Serial interface

USB

Parallel interface

IEEE 488, addresses 1 - 30

Analog output

0 - 10VDC to control an external
transformer

CN interface

15pin SubD connector to control an
external coupling network

Fail inputs

DUT monitoring via Fail1 and Fail2
input (one each)

DIMENSIONS

Dimensions

19", 3HU,
19", 6HU, (with TSurge5 module)

Weight

approx. 25kg

MAINS

Supply voltage

115V/230VAC +10%/-15%

Power

approx. 75W

Frequency

50/60Hz

Fuses

2 x T 2AT (230V) or 2 x T 4AT (115V)

SAFETY

Safety standard

EN/IEC 61010

Security circuit

Control input (24VDC)

Warning lamp

Floating contact (max. 230V/6A)

ACCESSORIES INCLUDED

Mains supply

Plug depends on the country of use

DUT supply

Plug depends on the country of use

DUT adapter

Socket depends on the country of
use

Operation manual,
Calibration certificate,
iec.control remote control software

OPTIONS

CNI 503Ax

3-phase coupling/decoupling
networks as per IEC 61000-4-4 and
-4-5 up to 100A per phase

iec.control 1

Remote control and documentation
software, including standard test
routines and reporting capabilities.

www.emtest.com

© EM TEST > PAGE 7/8