Combination wave / surge – Atec EM-Test-UCS500N5 User Manual

Page 5

background image

DATA SHEET > UCS 500N5 > 20120518

TECHNICAL DETAILS

COMBINATION WAVE / SURGE

SURGE MODULE, VCS/N5

As per EN/IEC 61000-4-5 and
EN 61000-6-1, -6-2

Voltage (o.c.)

160V - 5,000V ± 10%

Pulse front time

1.2us ± 30%

Pulse time to half
value

50us ± 20%

Current (s.c.)

Max. 2,500A ± 10%

Pulse front time

8us ± 20%

Pulse time to half
value

20us ± 20%

Polarity

Positive/negative/alternating

Event counter

1 - 30,000 or endless, selectable

TRIGGER CIRCUIT

Release of pulses

Automatic, manual, external

Synchronization

0° - 360°, resolution 1°

Repetition rate

max. 1Hz (1s - 999s)

OUTPUTS

Direct

Via HV connectors for external
coupling networks
(Zi = 2ohm with optional adapter
IMN 2)

Coupling mode

Line to line
Line(s) to ground

DUT supply

AC: 300V/16A; 50/60Hz
DC: 300V/10A

CRO trigger

5V trigger signal for oscilloscope

MEASUREMENTS

CRO Û-monitor

10Vp at 5,000V

CRO Î-monitor

10Vp at 2,500A

Peak voltage

5,000V in the LCD display

Peak current

2,500A in the LCD display

TEST ROUTINES

Quick Start

One-line adjustable parameters,
easy-to-use

Standard Test
routines

As per IEC 61000-4-5, Levels 1 - 4
As per EN 61000-6-1, -6-2
Manual Standard Test routine

User Test routines

Change polarity after n pulses
Change coupling after n pulses
Change voltage after n pulses
Change phase angle after n pulses

Pulsed Magnetic
Field

as per IEC 61000-4-9
Test levels 100, 300 and 1,000A/m
Test level steplessly adjustable
under Quick Start

OPTIONS

CNV504Nx

Coupling network for 4 signal/data
lines as per IEC 61000-4-5

CNV508Nx

Coupling network for 8 signal/data
lines as per IEC 61000-4-5

IMN 2

Impedance matching adapter to
match direct output to 2ohm source
impedance

www.emtest.com

© EM TEST > PAGE 5/8