DATA SHEET > UCS 500N5 > 20120518

TECHNICAL DETAILS

POWER FAIL, DIPS & INTER-
RUPTIONS, VOLTAGE VARIATIONS

POWER FAIL MODULE, PFS/N5

As per EN/IEC 61000-4-11,
IEC/EN 61000-4-29 and
EN 61000-6-1, -6-2

Channel PF1/PF2

AC voltage: max. 300V
AC current: max. 16A
DC voltage: max. 300V
DC current: max. 10A

Frequency

16Hz - 500Hz and DC

Switching time

< 5us into a 100ohm resistive load

Inrush current

> 500A

Protection

Both channels are protected against
short-circuit conditions.

TRIGGER CIRCUIT

Trigger of events

Automatic, manual, external

Synchronization

0° - 360°, resolution 1° (16 - 500Hz)

Repetition rate

10ms - 9,999s

Event duration

20us - 9,999s

OUTPUTS

DUT terminals

L, N and PE

CRO trigger

5V trigger signal for oscilloscope

MEASUREMENTS

DUT voltage

In the LCD display

DUT current

In the LCD display

MON V

Measurement of the DUT voltage;
built-in 100:1 divider

MON I

Measurement of the DUT current;
10mV/A; max. 1,000A

TEST ROUTINES

Quick Start

On-line adjustable parameters,
easy-to-use

Standard Test
routines

As per EN/IEC 61000-4-11
for AC supplies
As per EN/IEC 61000-4-29 for DC
supplies
As per EN 61000-6-1, -6-2
Manual Standard Test routine

User Test routines

Voltage variation, control of an
external variac
Change phase angle after n events
Change event duration after n
events
Inverse mode

50/60Hz magnetic
field

As per EN/IEC 61000-4-8
Test levels 1, 3, 10 and 30A/m with
external current transformer
MC2630
Test levels 100, 300 and 1,000A/m
with external current transformer
MC26100

OPTIONS

V4780

Tapped autotransformer as per IEC
61000-4-11 Ed.2

V4780 S2

Tapped autotransformer as per IEC
61000-4-11 Ed.2 with automatic
change of tap

MV2616

Motorised variac (0 - 250V, 16A)

MS100N

Magnetic field coil, 1m x 1m

MC2630

Current transformer for magnetic
fields up to 30A/m

MC26100

Current transformer for magnetic
fields up to 1,000A/m

CA PFS

Calibration box for inrush current
verification as per IEC 61000-4-11

www.emtest.com