Atec EM-Test-UCS500N5 User Manual
Page 3
DATA SHEET > UCS 500N5 > 20120518
TECHNICAL DETAILS
ACCESSORIES
V 4780 - TAPPED VOLTAGE TRANSFORMER FOR VOLTAGE
DIPS AND INTERRUPTIONS
The V 4780 tapped autotransformer is designed to supply
the required voltages as per IEC 61000-4-11 Ed.2:2004 to
perform voltage dips and interruptions.
V 4780S2 - TAPPED VOLTAGE TRANSFORMER FOR
VOLTAGE DIPS AND INTERRUPTIONS
The V 4780S2 tapped autotransformer is designed to
supply the required voltages as per IEC 61000-4-11
Ed.2:2004 to perform voltage dips and interruptions.
Compared to the manually operated V 4780 the V 4780S2
model offers automatic change of taps according to the
selected voltage level.
MS 100N - MAGNETIC FIELD COIL FOR POWER-FREQUENCY
AND PULSED MAGNETIC FIELDS
The MS 100N is a 1sqm magnetic field coil as specified in
IEC/EN 61000-4-8 and IEC/EN 61000-4-9. Its design
allows easy moving of the coil. The field coil is adjustable
in height and allows for 360degr rotation.
To generate power-frequency magnetic fields in the lower
range the current transformer MC 2630 is used while
high-field strength above 100A/m requires the MC 26100
current transformer.
HFK - CAPACITIVE COUPLING CLAMP
The HFK is a fully compliant capacitive coupling clamp as
per specification of IEC 61000-4-4.
ITP - IMMUNITY TEST PROBES
ITP is a tool being used for development test. It consists
of a variety of electrical field probes. The probes allow to
locate weak points within a system or on a PCB. The burst
pulse is used to generate the disturbance signal.
CA EFT KIT - VERIFICATION KIT FOR EFT/BURST PULSES
As per IEC 61000-4-4 Ed.2 the characteristic of the burst
generator needs to be verified with two different loads,
50ohm and 1,000ohm. EM TEST offers a calibration kit
consisting of the two loads and an adapter to verify the
pulses at the DUT output.
www.emtest.com
© EM TEST > PAGE 3/8