1. Frequency reference error = (aging rate x period of time since adjustment +

settability + temperature stability)

2. Not available in RBW < 1kHz
3. Marker level to DANL > 25 dB, Span

≤ 1.5 GHz, RBW/Span ≥ 0.002

4. N = LO harmonic mixing mode
5. Mixer power level (dBm) = input power (dBm) – input attenuator (dB)
6. Characteristic
7. Referenced to 50 MHz amplitude reference (20° C – 30° C)
8. Reference to midpoint between highest and lowest frequency response

deviations. (20° C – 30° C)

9. For reference levels 0 to 50 dBm; input attenuation 10 dB; dc coupled;

RFW 1 kHz; VBW 1 kHz; scale loge range 0 to -50 dB from reference level;
sweeptime coupled; signal input 0 to 50 dB; spsn

≤ 20 kHz.

10. Characteristic; factory preset, fixed center frequency, sweep points = 101

auto align off, RBW = 1 MHz, stop frequency

≤ 3 GHz, span > 10 MHz and

≤ 600 MHz (E4401, span > 102 MHz and ≤ 400 MHz).

11. Characteristic; factory preset, fixed center frequency, sweep points = 101 auto

align off, RBW = 1 MHz, stop frequency

≤ 3 GHz, span = 20 MHz, GPIB

interface, display and markers off, fixed center frequency, single sweep

12. In time domain sweeps
13. Reference level -25 dBm (E7401A) or -20 dBm (E7402A/03A/04A/05A);

input attenuation 10 dB; center frequency 50 MHz; RBW 1 kHz; VBW 1 kHz;
scale linear or log; span 2 kHz; sweep time coupled, sample director, signal at
reference level.

14. Meets Class A performance during dc operation or serial number

US41110000 or lower.

15. RBW

≥1 kHz, 2 sweep points

16. 10 Hz to 300 Hz are only available in spans of

≤ 5 MHz and are not usable

with tracking generator Option 1DN.

17. When storing a 401-point trace plus the instrument state