Atec Agilent-E7401A-02A-03A-04A-05A User Manual
Page 3
3
Maximum safe input level
Average continuous power
E7401A
(input attenuator
≥15 dB)
+30 dBm (1 W)
(input attenuator
≥5 dB)
E7402A/03A/04A/05A
+30 dBm (1 W)
Peak pulse power
(input attenuator
≥30 dB)
E7402A/03A/04A/05A
+50 dBm (100 W)
E7401A
+30 dBm (1 W)
dc
E7401A, E7402A
100 Vdc
E7402A (Option UKB)
0 Vdc (dc coupled)
50 V (ac coupled)
E7403A, E7404A
0 Vdc (dc coupled)
50 V (ac coupled)
E7405A
0 Vdc
(Option UKB)
0 Vdc (dc coupled)
50 V (ac coupled)
1 dB gain compression (total power at input mixer
5
)
≥ 50 MHz
0 dB
≥ 6.7 GHz
-3 dB
≥ 13.2 GHz
-5 dB
Displayed average noise level (dBm)
(Input terminated, 0 dB attenuation, sample-detector)
1 kHz RBW; 30 Hz VBW
10 Hz RBW; 1 Hz VBW
1 Hz RBW; 1 Hz VBW (Option 1D5)
1 kHz
10 Hz
1 kHz
10 Hz
1 Hz
RBW
RBW
w/preamp w/preamp Option 1D5
on
on, typical w/preamp
on, typical
E7401A
400 kHz to 10 MHz
≤ -115 ≤ -134 ≤ -150
≤ -155
≤ -165
10 MHz to 500 MHz
≤ -119 ≤ -138 ≤ -154
≤ -156
≤ -166
500 MHz to 1 GHz
≤ -117 ≤ -136 ≤ -152
≤ -156
≤ -166
1 GHz to 1.5 GHz
≤ -114 ≤ -133 ≤ -150
≤ -155
≤ -165
E7402A
30 Hz to 9 kHz
22
(Option UKB)
na
≤-93
na
na
na
9 kHz to 100kHz
22
na
≤-109
na
na
na
100 kHz to 1 MHz
22
na
≤-135
na
na
na
1 MHz to 10 MHz
22
≤-117 ≤-136
na
≤-152
≤-162
10 MHz to 1 GHz
≤-117 ≤-136
≤-152
19
≤-156
≤-166
1 GHz to 2 GHz
≤-116 ≤-135
≤–153
19
≤-156
≤-166
2 GHz to 3 GHz
≤-114 ≤-133
≤–151
19
≤-154
≤-164
E7403A, 04A, 05A
30 Hz to 9 kHz
22
(Option UKB)
na
≤ -93
na
na
na
9 kHz to 100kHz
22
na
≤ -109
6
na
na
na
100 kHz to 1 MHz
22
na
≤ -135
6
na
na
na
1 MHz to 10 MHz
22
≤ -117
6
≤ -137
6
na
≤ -155
≤ -165
10 MHz to 1 GHz
≤ -116 ≤ -135 ≤ -151
19
≤ -157
≤ -167
1 GHz to 2 GHz
≤ -116 ≤ -131 ≤ –151
19
≤ -155
≤ -165
2 GHz to 3 GHz
≤ -112 ≤ -131 ≤ –149
19
≤ -152
≤ -162
3 GHz to 6 GHz
≤ -112 ≤ -131 na
≤ -138
na
6 GHz to 12 GHz
≤ -111 ≤ -130 na
≤ -137
na
12 GHz to 22 GHz
≤ -107 ≤ -126 na
≤ -134
na
22 GHz to 26.5 GHz
≤ -106 ≤ -125 na
≤ -132
na
Display range
Log Scale
0.1, 0.2, 0.5 dB/division and 1 to 20 dB/division
in 1 dB steps; ten divisions displayed
RBW
≥ 1kHz
0 to -85 dB from reference level is calibrated
RBW
≤ 300 kHz 0 to -120 dB from reference level is calibrated
Linear scale
10 divisions
Scale units
dBm, dBmV, dBµV, dBµA, Amps, Volts and Watts
Marker readout resolution
Log scale
0 to -85 dB
0.04 dB
0 to -120 (RBW
≤ 300 Hz 0.04 dB
Linear scale
0.01% of reference level
Fast sweep times for zero span (Option AYX)
Log Scale
0 to -85 dB
0.3 dB
Linear
0.3 dB of reference level
Frequency response
(10 dB input attenuation)
Absolute
7
Typical
Relative
flatness
8
E7401A
9 kHz to 1.5 GHz
±0.5 dB
na
±0.5 dB
E7402A/03A/04A/05A
30 Hz to 3 GHz
6
(Option UKB)
±0.5 dB
na
±0.5 dB
9 kHz to 3 GHz
±0.46 dB
±0.14 dB
±0.5 dB
3.0 GHz to 6.7 GHz
±1.5 dB
±0.39 dB
±1.3 dB
6.7 GHz to 13.2 GHz
±2.0 dB
±0.68 dB
±1.8 dB
13.2 GHz to 26.5 GHz ±2.0 dB
±0.86 dB
±1.8 dB
Input attenuation switching uncertainty at 50 MHz
Attenuation setting
0 dB to 5 dB
±0.3 dB
10 dB
Reference
15 dB
±0.3 dB
20 to 60 dB (E7401A)
±(0.1 dB + 0.01 x attenuator setting)
20 to 65 dB
±(0.1 dB + 0.01 x attenuator setting)
Absolute amplitude accuracy
Typical
At reference settings
13
±0.34 dB
±0.13 dB
E7401A
±0.30 dB
±0.10 dB
Preamp on
16
±0.37 dB
±0.14 dB
Overall amplitude accuracy
9
±(0.54 dB + absolute
frequency response)