Atec Agilent-E7401A-02A-03A-04A-05A User Manual
Agilent e7400 a-series emc analyzers, Data sheet, Specifications
These specifications apply to the Agilent Technologies E7401A,
E7402A, E7403A, E7404A and E7405A EMC analyzers.
Specifications
All specifications apply over 0° C to +55° C unless otherwise
noted and are covered by the product warranty. The analyzer will
meet its specifications when: it’s within the one year calibration
cycle, AUTO ALIGN [ALL] is selected, stored a minimum 2 hours
within the operating temperature range, turned on for at least 5
minutes, and Align Now RF has been run once every 24 hour period.
Typical performance describes the level at which 80% of the units will
meet or exceed with a 95% confidence level over 20 to 30° C, but is
not covered in the product warranty. Characteristics describe
expected product performance levels that are not covered in the
product warranty.
Frequency specifications
Frequency range
E7401A
50
Ω
9 kHz to 1.5 GHz
E7402A
9 kHz to 3.0 GHz
dc coupled
30 Hz
6
to 3.0 GHz
ac coupled
100 kHz
6
to 3.0 GHz
E7403A
dc coupled
9 kHz to 6.7 GHz
dc coupled (Option UKB) 30 Hz
6
to 6.7 GHz
ac coupled
100 kHz to 6.7 GHz
Band
0
9 kHz to 3.0 GHz
1
2.85 GHz to 6.7 GHz
E7404A
dc coupled
9 kHz to 13.2 GHz
dc coupled (Option UKB) 30 Hz
6
to 13.2 GHz
ac coupled
100 kHz to 13.2 GHz
Band LO harmonic = N
0
1-
9 kHz to 3.0 GHz
(Option UKB)
30 Hz
6
to 3.0 GHz
1
1-
2.85 GHz to 6.7 GHz
2
2-
6.2 GHz to 13.2 GHz
E7405A
9 kHz to 26.5 GHz
Band LO harmonic = N
0
1-
9 kHz to 3.0 GHz
0
(Option UKB)
30 Hz
6
to 3.6 GHz
1
1-
2.85 GHz to 6.7 GHz
2
2-
6.2 GHz to 13.2 GHz
3
4-
12.8 GHz to 19.2 GHz
4
4-
18.7 GHz to 26.5 GHz
Frequency reference
(Option 1D5)
Aging
±2 x 10
-6
/year
±1 x 10
-7
/year
Temperature stability
±5 x 10
-6
±1 x 10
-8
Settability
±5 x 10
-7
±1 x 10
-8
Frequency readout accuracy
(start, stop, center, marker)
±(frequency indication
x frequency reference error
1
+ span accuracy + 15% of RBW
+ 10 Hz) + 1 Hz x N
4
Marker frequency counter
2
Accuracy
3
±(marker frequency x frequency
reference error
1
+ counter resolution)
Counter Resolution
Selectable from 1 Hz to 100 kHz
Frequency span
Range
0 Hz (zero span), 100 Hz x N
4
to
the range of the spectrum analyzer
Resolution
2 Hz x N
4
Accuracy (> 2000 sweep points)
Sweep type linear
±0.5% of span
Sweep type log
±2% of span (characteristic)
Agilent
E7400 A-series EMC Analyzers
Data Sheet