SPECIFICATIONS

15

Ground Continuity
For 75/20/30/40/50DT

Current : DC 0.1 A 0.01A, fixed

Max. ground resistance : 1

0.1 , fixed

Ground Fault Interrupt

GFI Trip Current:
HV Shut Down Speed:

450 A max (AC or DC)

< 1ms

INSULATION RESISTANCE TEST MODE Model 75/30/40/50DT, 7564SA

Output Voltage

Range:

100 - 1000 Volts DC

Resolution: 1 volt/step

Accuracy:

(2 of reading 2 volts)

Short Circuit Current

Maximum:

12mA peak

Voltage Display

Range:

0 - 1000 V

Resolution: 1 volt/step

Accuracy:

(2 of reading 2 counts)

Resistance Display

Range:

1 - 9999 M (4 Digit, Auto Ranging)

Resolution:

500VDC

1000VDC

M

M

M

0.001

1.000 - 5.388

1.000 - 9.999

0.01

1.40 - 53.88

2.80 - 99.99

0.1

14.0 - 538.8

28.0 - 999.9

1

104 - 9999

280 - 9999

Accuracy:

(2 of reading 2 counts) at test voltage

500 - 1000V and 1 - 1000 M

(8 of reading 2 counts) at test voltage

500 - 1000V and 1000 - 9999 M

(8 of reading 2 counts) at test voltage

100 - 500V and 0 - 1000 M

Charge-LO

Range:

0.000 - 3.500 A or Auto Set

HI-Limit

Range:

0 - 9999 M (0 = Off)

LO-Limit

Range:

1 - 9999 M

Delay Timer

Range:

0, 0.5 - 999.9 sec (0 = Constant)

Resolution: 0.1 sec/step

Accuracy:

(0.1% + 0.05 sec)

Ground Fault Interrupt

GFI Trip Current:
HV Shut Down Speed:

450 A max (AC or DC)

< 1ms