SPECIFICATIONS

14

Voltage Display

Range:

0.00 - 6.00 KV Full Scale
0.00 - 5.00 KV Full Scale for 7520DT only

Resolution: 10 volt/step

Accuracy:

(2 of reading 2 counts)

Current Display

Auto Range

AC Range 1:

0.000mA - 3.500mA

Resolution: 0.001mA/step

Range 2:

3.00 - 40.00 mA for 7550DT and 7564SA,
3.00 - 20.00 mA for 75/20/30/40DT

Resolution: 0.01 mA/step

DC Range 1:

0.0 A - 350.0 A

Resolution: 0.1 A/step

Range 2:

300 A - 3500 A

Resolution: 1 A/step

Range 3:

3000 A - 9990 A

Resolution: 10 A/step

Accuracy: All Ranges

(2 of reading 2 counts)

DC Output Ripple

4 Ripple RMS at 6 KV DC @ 3.5 mA, Resistive Load

Discharge Time

200 ms

Maximum Capacitive
Load DC Mode

1uF

< 1KV

0.75uF < 2KV
0.5uF < 3KV

0.08uF < 4KV
0.04uF < 5KV
0.01uF < 6KV

AC Output Wave Form Sine Wave, Crest Factor = 1.3 - 1.5

Output Frequency

Range:

60 or 50 Hz, User Selection

Accuracy:

1%

Output Regulation

(1 of setting 5 volts) from no load to full load

Dwell Timer

Range:

0, 0.3 - 999.9 sec (0 = Constant)

Resolution: 0.1 sec increments

Accuracy:

(0.1% + 0.05 sec)

Ramp Timer

Range:

AC 0.1 - 999.9 sec

DC 0.4 - 999.9 sec

Resolution: 0.1 sec increments

Accuracy:

(0.1% + 0.05 sec)