SPECIFICATIONS
13
Models 7520DT, 7530DT, 7540DT, 7550DT and 7564SA
Functional Specifications
INPUT
Voltage
115/230 VAC 15 , Single Phase, User selection
Frequency
50/60 Hz 5
Fuse
6.3 Amp 250V Slo-Blo for 7550DT and 7564SA
5 Amp 250V Slo-Blo for 75/20/30/40DT
DIELECTRIC WITHSTAND TEST MODE
Output Rating
5 KV @ 40 mA for 7550DT and 7564SA,
5 KV @ 20mA for 75/20/30/40DT
6 KV @ 10 mA DC for 75/30/40/50DT and 7564SA
Output Adjustment
Range:
0 - 5 KV AC
0 - 6 KV DC
Resolution: 1 volt/step
Accuracy:
(2 of setting 5 volts)
Ramp-HI
12mA peak maximum, ON/OFF selectable
Charge-LO
Range:
0.0 - 350.0 A DC or Auto set
HI-Limit AC Range:
0.00 - 40.00 mA for 7550DT and 7564SA,
0.00 - 20.00 mA for 75/20/30/40DT
Resolution: 0.01 mA/step
Accuracy:
(2 of setting 2 counts)
DC Range:
0 - 9999 A
Resolution: 1 A/step
Accuracy:
(2 of setting 2 counts)
LO-Limit AC Range:
0.000 - 9.999 mA
Resolution: 0.001 mA/step
Accuracy:
(2 of setting 2 counts)
DC Range:
0.0 - 999.9 A
Resolution: 0.1 A/step
Accuracy:
(2 of setting 2 counts)
Arc Detection
Range:
1 - 9
Failure Detector
Audible and Visual
Recycled Equipment (410)685-1997 www.recycledequipment.com email: [email protected]