6.
ITCU Universal thermocouple amplifier
Connector
Type
U
miniature
thermocouple
Isolation
250 VRMS or DC, Cat II.
Bandwidth
5 Hz update rate
Absolute Max Input
+/- 10V
Specified Range Type J:
-210 to 1200
°C
Specified Range Type K:
-200 to 1372
°C
Specified Range Type E:
-200 to 1000
°C
Specified Range Type T:
-200 to 400
°C
Specified Range Type N:
-200 to 1300
°C
Specified Range Type B:
600 to 1820
°C (250 to 1820 on menu)
Specified Range Type R:
-20 to 1768
°C
Specified Range Type S:
-20 to 1768
°C
A/D
24
bit
Sigma
Delta
Anti-Aliasing
Filter
Inherent
Resolution
0.01
°C
Thermocouple
types
J,K,E,T,N,B,R,S
Accuracy
(25°C)
J
+/-
0.8
°C
Accuracy
(25°C)
K
+/-
1.0
°C
Accuracy
(25°C)
E
+/-
0.8
°C
Accuracy (25°C) T (-100 to 400)
+/- 1.0
°C
Accuracy (25°C) T (-200 to -100)
+/- 2.0
°C
Accuracy
(25°C)
N
+/-
0.8
°C
Accuracy
(25°C)
B
+/-
2.0
°C
Accuracy
(25°C)
R
+/-
2.0
°C
Accuracy
(25°C)
S
+/-
2.0
°C
Accuracy (25°C) 100 mV
+/- 0.01% of attenuator
Cold Junction Compensation
Yes
Compensation Error
included in above accuracy specification
Intrinsic Noise (pk-pk, J,K,E,T,N)
< 0.07
°C
Intrinsic Noise (pk-pk, B,R,S)
< 0.5
°C
Intrinsic Noise (pk-pk, 100 mV)
< 0.005 % of attenuator
IMR
>
110
dB
@
DC
Linearization
NIST
ITS-90
Channels Per Module
1
Frequency
Counter
No
Module Part Number
32750-090
Note: Specified accuracy does not include probe errors.