IHV1 Isolated high voltage module


Connector

Guarded

banana

jacks

(red/black)

Isolation

250 VRMS or DC, Cat II. (iso-common to chassis
and other iso-commons)

Bandwidth (100V attenuator) 39 KHz (-3db)

Bandwidth (other attenuators)40 KHz (-3db)

Input

Single-ended,

DC

coupled

Zero

Suppression

Yes,

digital.

Measurement Ranges

+/- 400 V (400 VFS or 800 VFS w/ zero offset)

+/- 200 V (200 VFS or 400 VFS w/ zero offset)

+/- 100 V (100 VFS or 200 VFS w/ zero offset)

+/- 50 V (50 VFS or 100 VFS w/ zero offset)

+/- 40 V (40 VFS or 80 VFS w/ zero offset)

+/- 20 V (20 VFS or 40 VFS w/ zero offset)

+/- 10 V (10 VFS or 20 VFS w/ zero offset)

+/- 5 V (5 VFS or 10 VFS w/ zero offset)

Max Rated Input

+/- 250 Vrms or DC

Max Transient Input

+/- 800 V (not to exceed 250 Vrms)

A/D

16

bit

SAR

Anti-Aliasing Filter

4 pole Bessel

Accuracy (25°C)

+/- 0.15% of Attenuator

Cold Start Drift

Less than 0.2% of attenuator

Overshoot

Less than 2% of attenuator

Intrinsic Noise (pk-pk)

< 0.1 % of Attenuator

Min Input Impedance

1 Mohm

IMR at 60 Hz

> 75 dB

Channels Per Module

1

Frequency Counter

Yes. Software selectable.

Module Part Number

32750-010