3.
IBR1 Isolated differential/ bridge amplifier with DC excitation
Connector
5
wire
screw
terminal
Isolation
250 VRMS or DC, Cat II (iso-common to chassis
and other iso-commons)
Bandwidth
40
KHz
(-3db)
Input
Differential,
DC
coupled
Zero
Suppression
Yes,
digital
Absolute Max Input
+/- 12 V (either input referenced to iso-common)
Measurement Ranges
+/- 2 V
+/-
200
mV
+/- 50 mV
+/- 20 mV
Max Transient Input
60 V
A/D
16
bit
SAR
Anti-Aliasing Filter
4 pole Bessel
Accuracy (25°C)
+/- 0.15 % of attenuator
Cold Start Drift
Less than 0.2% of attenuator
Overshoot
Less than 2% of attenuator
Intrinsic Noise (pk-pk)
< 0.1 % of attenuator (2V, 200 mV Attenuators)
< 0.2 % of attenuator (50 mV Attenuator)
< 0.2 % of attenuator (20 mV Attenuator)
Min Input Impedance
250 K (balanced to signal common)
CMR at 60 Hz
> 90 dB
Excitation
DC adjustable to 10 V @ 30 mA
Excitation Range
0.1 to 10.0 V
Excitation
Accuracy
0.05
V
Auto Balance
yes (limited by maximum span)
Channels Per Module
1
Frequency Counter
Yes. Software selectable.
Bridge Completion Resistors: Yes, internal. Includes location for 3-wire ¼ bridge.
Module Part Number
32750-020
Specifications subject to change without notice