Measurement Accuracy

Conditions of accuracy specifications

Open/Short/50

calibration must be done. Calibration ON.

Averaging (on point) factor is larger than 32 at which calibration is done if Cal points is set to USER

DEF.

Measurement points are same as the calibration points.

Environment temperature is within ±5°C of temperature at which calibration is done, and within 13°C to

33°C. Beyond this environmental temperature condition, accuracy is twice as bad as specified.

7 mm connector is used.

When the analyzer frequency is identical to the transmitted interference signal frequency, refer to

"EMC" in "General Characteristics".

Contact Check Measurement Accuracy

@ measurement range : 0.1

to 100

, resolution : 1 m

± {3 + (25m

/R

dut

+ R

dut

/10k

)

ЧЧЧЧЧ

100} [%]

|Z|, |Y| Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ±(E

a

+ E

b

) [%]

θ

Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ± [rad]

L, C, X, B Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ±(E

a

+ E

b

)

×

[%]

R, G Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ±(E

a

+ E

b

)

×

[%]

D Accuracy (

D)

@|D

N

tan | < 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ±

Especially, @ D

x

0.1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ±

Q Accuracy (

Q)

@|Q

N

tan | < 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ±

Especially, @

Q

x

10 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ±Q

N

2

Where,

E

a

: depends on measurement frequency as follows:

1 MHz

Frequency

100 MHz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0.65 + [%]

100 MHz

500 MHz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0.8 + [%]

500 MHz

1000 MHz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.2 + [%]

E

b

: (Z

s

/|Z

x

| + Y

0

|Z

x

|)

ЧЧЧЧЧ

100 [%]

V

osc

: OSC level [ V ]

Z

x

: impedance measurement value [

]

Z

s

and Y

o

depend on number of point averaging (N

av

) and OSC level (V

osc

) as follows:

(E

=

+ E

>

)

100

(1 + D

N

)

(1 + Q

N

)

(E

=

+ E

>

)

100

(E

=

+ E

>

)

100

(E

=

+ E

>

)

100

10

(E

=

+ E

>

)

(E

=

+ E

>

)

100

0.03

8

OSC

0.03

8

OSC

0.03

8

OSC

(1 + D

N

2

)tan( )

1

µ

D

N

tan( )

E

=

+ E

>

100

E

=

+ E

>

100

(1 + Q

N

2

)tan( )

1

µ

Q

N

tan( )

E

=

+ E

>

100

E

=

+ E

>

100