Measurement function – Atec Agilent-4286A User Manual
Page 6
![background image](/manuals/318033/6/background.png)
6
Measurement Function
Measurement Parameters
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . L
p
-D, L
p
-Q, L
p
-G, L
p
-R
p
, L
s
-D, L
s
-Q, L
s
-R
s
, R-X, |Z|-
θ
rad
, |Z|-
θ
deg
,
C
p
-D, C
p
-Q, C
p
-G, C
p
-R
p
, C
s
-D, C
s
-Q, C
s
-R
s
, G-B, |Y|-
θ
rad
, |Y|-
θ
deg
Measurement Range
Impedance
@ 1MHz, accuracy < 10%, N
av
≥
8, V
osc
≥
0.2 V . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 200 m
Ω
to 3 k
Ω
Inductance
@ Q < 100, depends on frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 nH to 100
µ
H
Contact Check Function
Measurement Current . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . < 1 mA
List Sweep Characteristics
Sweep Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Continuous, Single, Manual
Sweep Direction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Up sweep
Number of Measurement Point . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 to 10 points
Averaging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Point average
Delay Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Point delay time, Sweep delay time