Typical Measurement Accuracy
Conditions of typical accuracy specifications
•
Open/Short/50
Ω
calibration must be done. Calibration ON.
•
Averaging (on point) factor is larger than 32 at which calibration is done if Cal points is set to USER
DEF.
•
Measurement points are same as the calibration points.
•
Environment temperature is within
±
5
°
C of temperature at which calibration is done, and within 13
°
C to
33
°
C. Beyond this environmental temperature condition, accuracy is twice as bad as specified.
•
7 mm connector is used.
|Z|, |Y| Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ±(E
a
+ E
b
) [%]
θ
Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ± [rad]
L, C, X, B Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ±(E
a
+ E
b
)
×
[%]
R, G Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ±(E
a
+ E
b
)
×
[%]
D Accuracy (
∆∆∆∆∆
D)
@ |D
N
tan | < 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ±
Especially, @ D
x
≤
0.1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ±
Q Accuracy (
∆∆∆∆∆
Q)
@ |Q
N
tan | < 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ±
Especially, @
≥
Q
x
≥
10 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ±Q
N
2
Where,
E
a
: depends on measurement frequency as follows:
1 MHz
≤
Frequency
≤
100 MHz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0.56 + 0.03 [%](Typical)
100 MHz ≤ 500 MHz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0.63 + 0.03 [%](Typical) 500 MHz ≤ 1000 MHz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0.70 + 0.03 [%](Typical) E b : (Z s /|Z x | + Y 0 |Z x |) ЧЧЧЧЧ 100 [%] V osc : OSC level [V] Z x : impedance measurement value [ Ω ] Z s and Y o depend on number of point averaging (N av ) and OSC level (V osc ) as follows: (E = + E > ) 100 N ) √ (E = + E > ) 100 (E = + E > ) 100 (E = + E > ) 100 E = + E > 100 (1 + Q N 2 )tan( ) E = + E > 100 1 µ Q N tan( ) 10 (E = + E > ) (E = + E > ) 100 0.03 8 OSC 0.03 8 OSC 0.03 8 OSC Measurement Conditions Z s [m Ω Ω Ω Ω Ω ] Y o [ µµµµµ S] Number of Point OSC Signal Level (Typical) (Typical) Averaging (N av ) (V osc ) 0.2 V ≤ V osc ≤ 1 V 20 + 0.05 × f [MHz] 7 + 0.1 × f [MHz] 0.01 V ≤ V osc < 0.2 V Ч (20 + 0.05 Ч f [MHz] ) Ч (7 + 0.1 Ч f [MHz] ) 0.2 V ≤ V osc ≤ 1 V 7 + 0.05 × f [MHz] 5 + 0.1 × f [MHz] 0.01 V ≤ V osc < 0.2 V Ч (7 + 0.05 Ч f [MHz] ) Ч (5 + 0.1 Ч f [MHz] ) 1 ≤ N av ≤ 7 N av ≥ 8 0.2 8 OSC 0.2 8 OSC 0.2 8 OSC 0.2 8 OSC (1 + D N 2 )tan( ) 1 µ D N tan( ) E = + E > 100 E = + E > 100 (1 + Q N ) √
(1 + D