Atec Agilent-34461A User Manual
Page 14
![background image](/manuals/318017/14/background.png)
14
34460A/34461A
G E N E R A L C H A R A C T E R I S T I C S
General characteristics
Line power
Power supply:
100/120 (127)/ 220 (230)/240 VAC ± 10%, CAT II
Power line frequency:
50/60/400 Hz ± 10%
Power consumption:
25 VA
Environment
Operating environment: Full accuracy for 0 to 55 °C
Full accuracy to 80% R.H. at 40 °C non–condensing
Operating altitude:
Up to 3000 m
Storage temperature:
-40 to 70 °C
Mechanical
Rack dimensions:
(W x H x D): 212.8 mm x 88.3 mm x 272.3 mm
Bench dimensions:
(W x H x D): 261.2 mm x 103.8 mm x 303.2 mm
Weight:
34460A: 3.68 kg (8.1 lb)
34461A: 3.76 kg (8.3 lb)
Triggering and memory
Samples per trigger
1 to 1,000,000
Trigger delay
0 to 3600 sec (~1 µs step size)
External trigger delay
<10 µs
External trigger jitter
<1 µs (DC fi xed range)
Volatile reading memory 10,000 (34461A), 1,000 (34460A)
Probe hold
Sensitivity fi xed at 1% of reading
Capture and navigate stable list of readings
Internal fl ash fi le system
80 MB total capacity
Save reading memory to non-volatile memory in CSV format
Store and recall user-defi ned states, power-off state,
1
and preference fi les
Save screen captures in BMP or PNG formats
Math functions
Per function null, min/max/avg/Sdev, dB, dBm, span, count, limit test, histogram
Display
4.3” color TFT WQVGA (480x272) with LED backlight
Supports: basic number, bar chart, trend chart (34461A only), histogram views
User-defi ned power-on message, display label, and selectable screen colors
Integrated, context-sensitive system help through press-and-hold buttons
Real-time clock/calendar
Set and read, year, month, day, hour, minute, seconds (Note: seconds not settable)
Battery CR-2032 coin-type, replaceable, >10-year life (typ)
Software available
IO Libraries: www.agilent.com/fi nd/IOLibraries
DMM Connectivity Utility software: www.agilent.com/fi nd/DMMutilitysoftware
System speeds (averages)
Benchmark
GPIB
USB 2.0
VXI-11
Sockets
Function change
1
50/s
50/s
50/s
50/s
Range change
2
100/s
100/s
100/s
100/s
1. Rate to change from 2-wire resistance to any other function
2. Rate to change from one range to the next higher range, ≤ 10 V, ≤ 10 MΩ
1. Power-off state only when power-down is initiated via front-panel power switch.
Regulatory
Safety
EN 61010-1:2010 (3rd Edition)
ANSI/ISA-61010-1 (82.02.01) Third Edition
ANSI/UL 61010-1 Third Edition
CAN/CSA-C22.2 No. 61010-1 Third Edition
EN 61010-2-030:2010 (1st Edition)
ANSI/ISA-61010-2-030 (82.02.03) First Edition
ANSI/UL 61010-2-030 First Edition
CAN/CSA-C22.2 No. 61010-2-030 First Edition
Refer to Declaration of Conformity for current revisions
Measurement Category II to 300 V
Other non MAINS circuits to 1000 Vpk
Pollution Degree 2
EMC
IEC 61326
EN 61326
CISPR
ICES-001
AS/NZS 2064.1
Refer to Declaration of Conformity for current revisions
Acoustic noise (nominal) 45 dBA
Triggering conditions
External input
Low–power TTL compatible input programmable edge triggered
Delay:
<1 µs
Jitter:
<1 µs
Minimum pulse width:
1 µs
Maximum rate:
Up to 1 kHz (34461A), up to 300 Hz (34460A)
Voltmeter complete output 3.3 V logic output
Polarity:
Programmable edge pulse
Pulse width:
Approximately 2 µs
Computer interfaces
LXI (rev 1.4)
10/100Base-T Ethernet (Sockets, VXI-11 protocol,
Web user interface) (Optional on 34460A)
USB
USB 2.0 (USB-TMC488 & MTP protocol)
GPIB
Optional GPIB IEEE-488
Language
SCPI-1999, IEEE-488.2, 34401A compatible
Front-panel USB host port
Supports USB 2.0 high-speed mass storage (MSC) class devices
Capability: import/export instrument confi guration fi les, save volatile readings
and screen captures