Performance versus measurement speed

For DC voltage, DC current, and resistance

1

34460A

34461A

Integration time

Digits

Readings/s

Digits

Readings/s

Additional noise error

100 PLC/1.67 s (2 s)

0.6 (0.5)

0.6 (0.5)

0% of range

10 PLC/167 ms (200 ms)

6 (5)

6 (5)

0% of range

1 PLC/16.7 ms (20 ms)

60 (50)

60 (50)

0.001% of range

0.2 PLC/3 ms (3 ms)

100

300

0.001% of range

2

0.02 PLC/400 µs (400 µs)

300

1000

0.01% of range

2

AC voltage, AC current

3, 4

Digits

ACV

ACI

AC fi lter

.4/s

.6/s

Slow

1.6/s

4/s

Medium

40/s

40/s

Fast

50/s

5

50/s

5

Fast

Frequency, period

Aperture

Digits

Readings

1 Second

7

1

0.1 Second

6

10

0.01 Second

5

80

System Speeds (average)

DC voltage, DC current, resistance

1, 2

34460A

34461A

Autorange time

3

<30 ms

<30 ms

Maximum internal trigger rate

300/s

1000/s

Maximum external trigger rate

300/s

1000/s

ASCII readings to bus

300/s

1000/s

Single reading transaction rate

4

50/s

150/s

AC voltage, AC current

5

Autorange time

3

10/s

10/s

Maximum internal trigger rate

50/s

50/s

Maximum external trigger rate

50/s

50/s

ASCII readings to bus

50/s

50/s

Single reading transaction rate

4

50/s

50/s

5

Frequency, period

6

Autorange time

3

10/s

10/s

Maximum internal trigger rate

80/s

80/s

Maximum external trigger rate

80/s

80/s

ASCII readings to bus

80/s

80/s

Single reading transaction rate

4

50/s

50/s

34460A/34461A

O P E R AT I N G C H A R A C T E R I S T I C S

1. Reading speeds for 60 Hz and 50 Hz operation, autozero off, fi xed range.

2. Add 20 µV for DCV and 20 mΩ for resistance.

Add 0.2 µA for DC current + 10x the above range error for the 10 mA range.

3. Maximum reading rates for 0.01% of AC step additional error.

Additional settling delay required when input DC level varies.

4. For external trigger or remote operation using default settling delay (Delay Auto).

5. Maximum useful limit with default settling delays defeated.

1. 0.02 NPLC, delay 0, autozero off, math off, and display off.

2. These rates apply to all I/O interfaces.

3. Time to automatically change one range and be ready for new measurement, ≤10 V, ≤10 MΩ.

4. Includes measurement and IO time (assumes connection via SOCKETS. VXI-11 connections may be slower).

5. Fast AC fi lter, delay 0, math off, and display off.

6. 10-ms aperture, fast AC fi lter, delay 0, math off, and display off.

34460A DMM rear panel with GPIB option installed.

34461A DMM rear panel with GPIB option installed.