Atec Agilent-E5260A User Manual

Page 5

background image

5

Output terminal/connection:
Triaxial connector, Kelvin (remote
sensing)

Voltage/current
compliance(limiting)
The SMU can limit output voltage
or current to prevent damaging the
device under test.
Voltage: 0 V to ± 200 V
Current: ± 100 pA to ± 1 A
Compliance accuracy: Same as the
current (or voltage) set accuracy.

HPSMU supplemental information
Maximum allowable cable resistance
(Kelvin connection):
Force line: 10 Ω (I ≤ 100 mA)
Force line: 1.5 Ω (100 mA < I ≤ 1 A)
Sense line: 10 Ω (All cases)
Voltage source output resistance:
0.2 Ω typical (Force line, non-Kelvin

connection)

Power consumption (HPSMU)
Voltage source mode:

Voltage range

Power

2 V

20 x Ic (W)

20 V

20 x Ic (W)

40 V

40 x Ic (W)

100 V

100 x Ic (W)

200 V

200 x Ic (W)

Where Ic is the current compliance setting.

Current source mode:

Voltage compliance

Power

Vc ≤ 20

20 x Io (W)

20 < Vc ≤ 40

40 x Io (W)

40 < Vc ≤ 100

100 x Io (W)

100 < Vc ≤ 200

200 x Io (W)

Where Vc is the voltage compliance setting
and Io is output current.

HPSMU measurement and output range

Voltage measurement input resistance:

10

13

Ω

Current source output resistance:

10

13

Ω (1 nA range)

Current compliance setting accuracy
(for opposite polarity):
For 1 nA to 10 nA ranges:
I setting accuracy ± 12 % of range
For 100 nA to 1 A ranges:
I setting accuracy ± 2.5 % of range
Maximum capacitive load:
For 100 nA to 10 mA ranges: 10 nF
For 100 mA to 1 A ranges: 100 µF
Maximum guard capacitance: 900 pF
Maximum shield capacitance: 5000 pF
Maximum guard offset voltage: ± 1 mV
Noise characteristics (typical, filter ON):
Voltage source: 0.01 % of V range

(rms)

Current source: 0.1 % of I range (rms)
Overshoot (typical, filter ON):
Voltage source: 0.03 % of V range
Current source: 1 % of I range

Range switching transient noise
(typical, filter ON):
Voltage ranging: 250 mV
Current ranging: 10 mV
Slew rate: 0.2 V/µs
SMU pulse setting accuracy
(fixed measurement range):
Width: 0.5 % + 50 µs
Period: 0.5 % + 100 µs
Trigger out delay (pulsed
measurements):
0 to 32.7 ms with 100 µs resolution

(< pulse width)

1000

500

125

50

-50

-125

-500

-1000

-200 -100 -40 -20

20 40 100 200

Current (mA)

Voltage (V)