Hpsmu (high power smu) module specifications – Atec Agilent-E5260A User Manual

Page 4

background image

4

Output terminal/connection:
Triaxial connector, Kelvin (remote
sensing)

Voltage/current compliance (limiting)
The SMU can limit output voltage
or current to prevent damaging the
device under test.
Voltage: 0 V to ± 100 V
Current: ± 100 pA to ± 200 mA
Compliance accuracy: Same as the
current (or voltage) set accuracy.

MPSMU supplemental information
Maximum allowable cable resistance
(Kelvin connection):
Force Line: 10 Ω (I ≤ 100 mA)
Force Line: 1.5 Ω

(100 mA < I ≤ 200 mA)

Sense Line: 10 Ω (All cases)

Voltage source output resistance:
0.3 Ω typical (Force line, non-Kelvin

connection)

Voltage measurement input resistance:

≥ 10

13

Ω

Current source output resistance:

≥ 10

13

Ω (1 nA range)

Current compliance setting accuracy
(for opposite polarity):
For 100 nA to 200 mA ranges:

I setting accuracy ± 2.5 % of range

Maximum capacitive load:
For 100 nA to 10 mA ranges: 10 nF
For 100 mA to 200 mA ranges: 100 µF
Maximum guard capacitance: 900 pF
Maximum shield capacitance: 5000 pF
Maximum guard offset voltage: ± 1 mV
Noise characteristics (typical, filter ON):
Voltage

source:

0.01 % of V range (rms)

Current source: 0.1 % of I range (rms)

Overshoot (typical, filter ON):

Voltage source: 0.03 % of V range

Current source: 1 % of I range
Range switching transient noise
(typical, filter ON):
Voltage ranging: 250 mV

Current ranging: 10 mV

Slew rate: 0.2 V/µs
SMU pulse setting accuracy
(fixed measurement range):
Width: 0.5 % + 50 µs
Period: 0.5 % + 100 µs
Trigger out delay (pulsed
measurements):
0 to 32.7 ms with 100 µs resolution

(< pulse width)

HPSMU (High Power SMU) Module Specifications

Voltage range, resolution, and accuracy (HPSMU)

Voltage
range

Force
resolution

Measure resolution

Force accuracy

1

Measure

accuracy

1

Maximum

current

±2 V

100 µV

100 µV

±(0.03 % + 900 µV)

±(0.03 % + 700 µV)

1 A

±20 V

1 mV

1 mV

±(0.03 % + 4 mV)

±(0.03 % + 4 mV)

1 A

±40 V

2 mV

2 mV

±(0.03 % + 7 mV)

±(0.03 % + 8 mV)

2

±100 V

5 mV

5 mV

±(0.04 % + 15 mV)

±(0.03 % + 20 mV)

3

±200 V

10 mV

10 mV

±(0.045 % + 30 mV)

±(0.035 % + 40 mV)

4

1. ± (% of output/measured value + offset voltage V)
2. 1 A (Vo ≤ 20 V), 500 mA (20 V 3. 1 A (Vo ≤ 20 V), 500 mA (20 V 4. 1 A (Vo ≤ 20 V), 500 mA (20 V

Current range, resolution, and accuracy (HPSMU)

Current
range

Force
resolution

Measure
resolution

Force accuracy

1,2

Measure accuracy

1,2

Maximum

voltage

±100 nA

5 pA

5 pA

±(0.12 % + 50 pA + 5 pA x (Vo/25))

±(0.1 % + 30 pA + 5 pA x (Vo/25))

200 V

±1 µA

50 pA

50 pA

±(0.12 % + 400 pA + 50 pA x (Vo/25))

±(0.1 % + 200 pA + 50 pA x (Vo/25))

200 V

±10 µA

500 pA

500 pA

±(0.12 % + 5 nA + 500 pA x (Vo/25))

±(0.1 % + 3 nA + 500 pA x (Vo/25))

200 V

±100 µA

5 nA

5 nA

±(0.12 % + 40 nA + 5 nA x (Vo/25))

±(0.1 % + 20 nA + 5 nA x (Vo/25))

200 V

±1 mA

50 nA

50 nA

±(0.12 % + 500 nA + 50 nA x (Vo/25))

±(0.1 % + 300 nA + 50 nA x (Vo/25))

200 V

±10 mA

500 nA

500 nA

±(0.12 % + 4 µA + 500 nA x (Vo/25))

±(0.1 % + 2 µA + 500 nA x (Vo/25))

200 V

±100 mA

5 µA

5 µA

±(0.12 % + 50 µA + 5 µA x (Vo/25))

±(0.1 % + 30 µA + 5 µA x (Vo/25))

3

±1 A

50 µA

50 µA

±(0.5 % + 500 µA + 50 µA x (Vo/25))

±(0.5 % + 300 µA + 50 µA x (Vo/25))

4

1. ± (% of output/measured value + offset current A (fixed part determined by the output/measurement range + proportional part that is multiplied by Vo)
2. Round up below decimal point for the value of (Vo/25).
3. 200 V (Io ≤ 50 mA), 100 V (50 mA < Io ≤ 100 mA)
4. 200 V (Io ≤ 50 mA), 100 V (50 mA < Io ≤ 125 mA), 40 V (125 mA < Io ≤ 500 mA), 20 V (500 mA < Io ≤ 1 A), Io is the output current in amps.