Atec Agilent-8920B User Manual
Page 9

9
Option 051, Rear Panel Connectors for TDMA
Measurements for TDMA Option 800/801
This option provides the necessary rear panel connectors to work
with the 83206A TDMA cellular adapter (8920B Option 800/801).
Option 055, Mechanical Variable Input Attenuator
This option replaces the standard electronic input attenuator with
a mechanical version. Option 055 extends the RF source frequency
range to 250 kHz to 1 GHz, and the RF analyzer measurement
range to 400 kHz to 1 GHz. This option can be ordered with all
options except Options 006 and 007.
Option 102, Spectrum Analyzer with Tracking
Generator and ACP
Frequency Range:
Standard: 10 MHz to 1 GHz
Option 055 (mechanical variable attenuator): 400 kHz to 1 GHz
Frequency Span/Resolution Bandwidth (coupled):
Span
Bandwidth
<50 kHz
300 Hz
<200 kHz
1 kHz
<1.5 MHz
3 kHz
<18 MHz
30 kHz
≥
18 MHz
300 kHz, plus full span capability
Display: Log with 1, 2, and 10 dB/div.
Display Range: 80 dB
Reference Level Range: +50 to –50 dBm
Residual Responses: <–70 dBm (no input signal, 0 dB attenuation
for reference level
≤
–20 dBm)
Image Rejection: >50 dBm for reference level
≤
30 dBm
Non-harmonic Spurious Responses: >60 dBc (for input signals
≤
–30 dBm at ANT IN port)
Supplemental Characteristics
Level Accuracy: ±2.5 dB
ANT IN Displayed Average Noise Level:
<–114 dBm for <50 kHz spans at reference level of –50 dBm
Log Scale Linearity: ±2 dB (for input levels
≤
–50 dBm and/or
60 dB range)
Tracking Generator (In Option 102)
Frequency Range:
Standard and Options 006, 007, and 016: 10 MHz to 1 GHz
Option 055 (mechanical variable attenuator): 400 kHz to 1 GHz
Frequency Offset: Frequency span endpoints ± frequency offset
cannot be <400 kHz or
≥
1 GHz
Output Level Range: Same as signal generator
Sweep Modes: Normal and inverted
Adjacent Channel Power (in Option 102)
Relative Measurements:
Level Range:
Antenna In: –40 dBm to +20 dBm
RF IN/OUT:
Standard: 1 mW to 60 W continuous or to 100 W for
10 seconds per minute
Option 006 (cellular power option): 50 µW to 10 W contin-
uous, 7.5 W for 10 seconds per minute
Option 007 (low power option): 40 µW to 2.4 W continuous,
2.0 W for 10 seconds per minute
Option 016 (high power option): 1.6 mW to 100 W
continuous, 125 W for 10 seconds per minute
Option 055:
Level Range: –137 to –19 dBm into 50 W
Reverse Power: 60 watts continuous, 100 watts for
10 seconds per minute.
W/Option 006 (cellular power option):
Level Range: –137 to –12.5 dBm into 50 W
Reverse Power: 5.0 watts continuous, 7.5 watts for
10 seconds per minute.
W/Option 007 (low power option):
Level Range: –137 to –6.5 dBm into 50 W
Reverse Power: 1.2 watts continuous, 2.0 watts for
10 seconds per minute.
W/Option 016 (high power option):
Level Range: –137 to –21 dBm into 50 W
Reverse Power: 100 watts continuous,
125 watts for 10 seconds per minute.
Dynamic Range: Typical values for channel offsets
Channel offset Channel BW
Dynamic range
12.5 kHz
8.5 kHz
–65 dBc
20 kHz
14 kHz
–68 dBc
25 kHz
16 kHz
–68 dBc
30 kHz
16 kHz
–68 dBc
60 kHz
30 kHz
–65 dBc
Relative Accuracy: Typically ±2.0 dB