beautypg.com

Atec Agilent-8920B User Manual

Page 7

background image

7

Residual Noise: 150 µV (15 kHz BW), <136 µV typical

Supplemental Characteristics
3 dB Bandwidth:
Typically 2 Hz to 100 kHz
Nominal Input Impedance: Switchable between 1 M

in parallel

with 95 pF or 600

floating

Resolution: For inputs

100 mV; 0.1 mV for inputs <100 mV; 0.01 mV

DC Voltage Measurement

Voltage Range: 100 mV to 42 V

Accuracy: ±1.0% of reading plus ±45 mV

Supplemental Characteristics
Resolution:
1 mV

Distortion Measurement

Tone Frequency Range: 300 Hz to 10 kHz

Tone Accuracy: ±5%

Input Level Range: 30 mV to 30 Vrms

Display Range: 0.1% to 100%

Accuracy: ±1 dB (0.5 to 100% distortion) for tones from 300 to
1500 Hz measured with the 15 kHz LPF.

±1.5 dB (1.5 to 100% distortion) for tones from 300 Hz to 10 kHz
measured with >99 kHz LPF.

Residual THD+Noise: –60 dBc or 150 µV, whichever is greater, for
tones from 300 to 1500 Hz measured with the 15 kHz LPF.

–52 dBc or 450 µV, whichever is greater, for tones from 300 Hz to
10 kHz measured with >99 kHz LPF.

Supplemental Characteristics
Resolution:
0.1% Distortion

SINAD Measurement

Tone Frequency Range: 300 Hz to 10 kHz

Tone Accuracy: ±5%

Input Level Range: 30 mV to 30 Vrms

Display Range: 0 to 60 dB

Accuracy: ±1 dB (0 to 46 dB SINAD) for tones from 300 to 1500 Hz
measured with the 15 kHz LPF.

±1.5 dB (0 to 36 dB SINAD) for tones from 300 Hz to 10 kHz meas-
ured with >99 kHz LPF.

Residual THD+Noise: –60 dBc or 150µV, whichever is greater, for
tones from 300 to 1500 Hz measured with the 15 kHz LPF.

–52 dBc or 450 µV, whichever is greater, for tones from 300 Hz to
10 kHz measured with >99 kHz LPF.

Supplemental Characteristics
Resolution:
0.01 dB

Audio Filters

Standard: <20 Hz HPF, 50 Hz HPF, 300 Hz HPF, 300 Hz LPF, 3 kHz
LPF, 15 kHz LPF, >99 kHz LPF, 750 µs de-emphasis

Optional: C-Message, CCITT, 400 Hz HPF, 4 kHz BPF, 6 kHz BPF

Variable Frequency Notch Filter

Frequency Range: 300 Hz to 10 kHz

Notch Depth: >60 dB

Notch Width: Typically ±5%

Audio Detectors

RMS, RMS*SQRT2, Pk+, Pk–, Pk+hold, Pk–hold,
Pk±/2, Pk±/2 hold, Pk±max, Pk±maxhold

Oscilloscope Specifications

Frequency Range: 2 Hz to 50 kHz (3 dB BW)

Amplitude Accuracy: ±1.5% of reading ±0.1 div. (20 Hz to 10 kHz)

Scale/Division: 10 mV to 10 V, settable 20 µV to 20 V

Time/Division: 10 µs to 100 ms, settable

Trigger Delay Range: 20 µs to 3.2 seconds

Supplemental Characteristics
3 dB Bandwidth:
Typically >100 kHz
Internal DC Offset:

0.1 div. (

50 µV/div. sensitivity)

Standard User Memory, RAM

Approximately 928 KB of RAM are available for nonvolatile
save/recall of settings. This typically will allow you to save
greater than 500 sets of instrument settings, depending on the
type of information saved.

Standard Rear Panel Interfaces

Current Sensing and I/O: GPIB/RS-232/Parallel (Centronics)

DC Current Meter

Measurement Range: 0 to 15 A (usable to 20 A)

Accuracy: The greater of ±10% of reading after zeroing or ±30 mA
(levels >100 mA)

Remote Programming

GPIB: A general purpose implementation of IEEE Standard 488.2