Physio-Control LIFEPAK CR Plus User Manual
Page 76
![background image](/manuals/313531/76/background.png)
D-4
LIFEPAK CR Plus and LIFEPAK EXPRESS Defibrillator Operating Instructions
Electromagnetic Compatibility Guidance
Table D-4
Recommended Separation Distances between Portable and Mobile RF Communications Equipment and the
LIFEPAK CR Plus and LIFEPAK EXPRESS defibrillators
Recommended Separation Distances between Portable and Mobile RF Communications
Equipment and the LIFEPAK CR Plus and LIFEPAK EXPRESS defibrillators
The LIFEPAK CR Plus and LIFEPAK EXPRESS defibrillators are intended for use in an electromagnetic
environment in which radiated RF disturbances are controlled. The customer or the user of the defibrillator can
help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the defibrillator as recommended below, according to the
maximum output power of the communications equipment.
Rated maximum
output power of
transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz 80 MHz to 800 MHz
80 MHz to 870 MHz,
900 MHz to 2.5 GHz
870 MHz to 900 GHz
0.01
Not Applicable
0.12
0.23
0.77
0.1
Not Applicable
0.38
0.73
2.43
1
Not Applicable
1.2
2.3
7.7
10
Not Applicable
3.8
7.3
24.3
100
Not Applicable
12
23
77
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in
meters (m) can be determined using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note 1:
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2:
The ISM (industrial, scientific and medical) bands between 150 kHz and 80 MHz are 6,765 MHz to
6,795 MHz; 13,553 MHz to 13,567 MHz; 26,957 MHz to 27,283 MHz; and 40,66 MHz to 40,70 MHz.
Note 3:
An additional factor of 10/3 is used in calculating the recommended separation distance for
transmitters in the ISM frequency bands between 150 kHz and 80 MHz and in the frequency range 80 MHz to
2.5 GHz to decrease the likelihood that mobile/portable communications equipment could cause interference if
it is inadvertently brought into patient areas.
Note 4:
These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
d
1.2 P
=
d
2.3 P
=
d
7.7 P
=