Physio-Control LIFEPAK CR Plus User Manual
Page 41
![background image](/manuals/313531/41/background.png)
LIFEPAK CR Plus and LIFEPAK EXPRESS Defibrillator Operating Instructions
5-3
©2009-2010 Physio-Control, Inc.
5
C
ari
ng f
o
r
Y
o
u
r Defi
bril
lator
Caring for Your Defibrillator
The replacement kit includes a CHARGE-PAK battery charger, one or two QUIK-PAK electrode packets,
and a CHARGE-PAK battery discharger. The discharger depletes a used battery charger so that it is
ready for recycling or disposal.
Note:
Always keep the CHARGE-PAK battery charger with its original defibrillator. When a
CHARGE-PAK battery charger is inserted into a defibrillator, it becomes internally linked to that
device. If the CHARGE-PAK battery charger is inserted into a second device, the device will not
function correctly. If this occurs, obtain a new replacement kit and replace both the CHARGE-PAK
battery charger and the QUIK-PAK electrodes at the same time.
Follow instructions provided in the replacement kit for battery charger and electrode packet
recycling/disposal.
WARNING!
Possible explosion or fire.
The CHARGE-PAK battery charger is not rechargeable. Do not attempt to recharge, open, crush, or
burn the battery, or it may explode or catch fire.