Physio-Control LIFEPAK CR Plus User Manual
Page 54
![background image](/manuals/313531/54/background.png)
A-2
LIFEPAK CR Plus and LIFEPAK EXPRESS Defibrillator Operating Instructions
Specifications
Waveform Parameters:
Note: Table values are nominal for a 360 joule energy setting.
Figure A-2
Biphasic Waveform
Patient
Impedance (
)
I1 (A)
I2 (A)
I3 (A)
I4 (A)
T1 (ms) T2 (ms)
25
69.5
28.3
28.3
15.5
5.6
3.7
50
37.9
19.8
19.8
12.8
7.4
4.9
75
26.1
15.6
15.6
11.0
8.5
5.7
100
19.9
12.7
12.7
9.4
9.7
6.5
125
16.0
10.9
10.9
8.4
10.4
6.9
150
13.5
9.5
9.5
7.6
11.0
7.4
175
11.6
8.5
8.5
6.9
11.6
7.7
Shock Advisory System: An ECG analysis system that advises whether a shock is appropriate, meets
rhythm recognition criteria specified in the American Association of Medical
Instrumentation standard DF39.
The device allows a defibrillation shock only if the Shock Advisory System
advises defibrillation.
Device Capacity:
LIFEPAK CR Plus defibrillator — Thirty (30) full discharges or 210 minutes
of “ON time” with a fully charged device.
LIFEPAK EXPRESS defibrillator — Twenty (20) full discharges or
140 minutes of “ON time” with a fully charged device.
T1
T2
I1
I2
I3
I4